Interference of hemoglobin (Hb) Las Palmas with KPLC measurement of HbA_(1c) in 87 patients

机译:血红蛋白(HB)LAS PALMAS在87名患者中具有KPLC测量的HBA_(1C)的干扰

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Hemoglobin A_(1c) (HbA_(1c)) can be measured using different testing modalities and, depending on the method applied, several factors may interfere in its measurement, which may lead to falsely abnormal results (1-3). Between November 2009 and January 2011, HbA_(1c) blood samples of 22,371 diabetic patients were analyzed in our institution. During this period, we detected an interference in the measurements of HbA_(1c), using a HPLC ADAMS A_(1c) HA-8160 of A. MENARINI Diagnostics method, in 87 diabetic patients (51 women and 26 men) living in the island Gran Canaria (Canary Islands, Spain). HbA_(1c) whole blood measurements were performed in EDTA K3 tubes, utilizing the HPLC HA-8160 method in Diabetic mode, and calibrated in accordance with the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Reference Measurement Procedure.
机译:血红蛋白A_(1C)(HBA_(1C))可以使用不同的测试模式测量,并且根据所施加的方法,若干因子可能会干扰其测量,这可能导致错误异常结果(1-3)。 在2009年11月至2011年1月期间,在我们的机构分析了22,371名糖尿病患者的HBA_(1C)血液样本。 在此期间,我们检测到HBA_(1C)测量的干扰,使用A. menarini诊断方法的HPLC ADAMS A_(1C)HA-8160,在岛上的87名糖尿病患者(51名女性和26名男子)中 大加那利岛(加那利群岛,西班牙)。 HBA_(1C)全血液测量在EDTA K3管中进行,利用糖尿病模式的HPLC HA-8160方法,并根据国际临床化学和实验室医学联合会(IFCC)参考测量程序校准。