Interference of hemoglobin (Hb) Las Palmas with KPLC measurement of HbA_(1c) in 87 patients

机译:KPLC对HbA_(1c)的KPLC测量对Las Palmas血红蛋白(Hb)的干扰

获取原文
获取原文并翻译 | 示例

摘要

Hemoglobin A_(1c) (HbA_(1c)) can be measured using different testing modalities and, depending on the method applied, several factors may interfere in its measurement, which may lead to falsely abnormal results (1-3). Between November 2009 and January 2011, HbA_(1c) blood samples of 22,371 diabetic patients were analyzed in our institution. During this period, we detected an interference in the measurements of HbA_(1c), using a HPLC ADAMS A_(1c) HA-8160 of A. MENARINI Diagnostics method, in 87 diabetic patients (51 women and 26 men) living in the island Gran Canaria (Canary Islands, Spain). HbA_(1c) whole blood measurements were performed in EDTA K3 tubes, utilizing the HPLC HA-8160 method in Diabetic mode, and calibrated in accordance with the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Reference Measurement Procedure.
机译:血红蛋白A_(1c)(HbA_(1c))可以使用不同的测试方式进行测量,并且根据所应用的方法,可能会因多种因素而影响其测量,从而可能导致错误的异常结果(1-3)。在2009年11月至2011年1月之间,我们对22,371名糖尿病患者的HbA_(1c)血液样本进行了分析。在此期间,我们使用A.MENARINI Diagnostics方法的HPLC ADAMS A_(1c)HA-8160检测了HbA_(1c)的测量结果,该方法对居住在岛上的87位糖尿病患者(51位女性和26位男性)大加那利岛(西班牙加那利群岛)。 HbA_(1c)全血测量是在EDTA K3试管中,以HPLC模式使用HPLC HA-8160方法进行的,并根据国际临床化学和实验室医学联合会(IFCC)参考测量程序进行了校准。