首页> 外文期刊>Clinical Biochemistry >Evaluation of white blood cell count in peritoneal fluid with five different hemocytometers
【24h】

Evaluation of white blood cell count in peritoneal fluid with five different hemocytometers

机译:五种不同血细胞计的腹膜液中白细胞计数评价

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives: Evaluation of automated flow cytometric analysis of white blood cell (WBC) count in peritoneal fluids. Methods: One hundred peritoneal fluids were analyzed with manual microscopy, Sysmex XE-2100 and XE-5000, Siemens Advia 2120, Mindray BC-6800, Abbott Sapphire. Results: High correlations (0.978 to 0.999) and modes biases (-132 to 80 WBC/mm3) were found. Agreement at septic peritonitis cutoff ranged between 96% and 99%. Conclusions: These hemocytometers display acceptable performance for WBC screening in peritoneal fluids.
机译:目的:评价腹膜液中白细胞(WBC)计数的自动流动细胞络分析。 方法:用手动显微镜,Sysmex XE-2100和XE-5000,西门子Advia 2120,Mindray BC-6800,Abbott Sapphire,分析了一百个腹膜液。 结果:找到高相关(0.978至0.999)和模式偏差(-132至80 WBC / mm3)。 脓毒症腹膜炎的协议范围在96%和99%之间。 结论:这些血细胞计数仪在腹膜液中显示WBC筛选的可接受性能。