首页> 外文期刊>Circulation. Arrhythmia and electrophysiology >Feasibility of Rapid Linear-Endocardial and Epicardial Ventricular Ablation Using an Irrigated Multipolar Radiofrequency Ablation Catheter
【24h】

Feasibility of Rapid Linear-Endocardial and Epicardial Ventricular Ablation Using an Irrigated Multipolar Radiofrequency Ablation Catheter

机译:利用灌溉多极射频消融导管的快速线性 - 内膜和心外膜心室消融的可行性

获取原文
获取原文并翻译 | 示例
           

摘要

Background-A common strategy for ablation of scar-based ventricular tachycardia is delivering multiple lesions in a linear pattern.Methods and Results-We tested the efficacy of a novel linear irrigated multipolar ablation catheter capable of creating linear lesions with a single application. Healthy swine underwent endocardial and epicardial linear ablation using a novel linear irrigated ablation catheter; control animals underwent focal lesions in a linear pattern over 3.5 cm with an irrigated radiofrequency catheter. The linear catheter contained 7 irrigated electrodes spaced over 3.5 cm and could deliver ^25 W to each electrode. Linear ablation required significantly less radiofrequency time than focal ablation (56±11 versus 497±110 seconds; P<0.0001). At gross pathology, linear (n=18) epicardial lines were longer than focal (n=8) epicardial lines (3.3±0.7 versus 2.1±0.9 cm; P<0.0005), with greater volume (3.8±2.9 versus 1.5±1.6 cm~3; P=0.002). There was no difference between linear (n=22) and focal (n=7) endocardial line length or volume. Gaps (length 2.8±0.9 mm) were present in 53% of focal lines and 0% of linear ablation lines. No perforations, steam pops, or thrombus were noted. Conclusions-Compared with sequential focal radiofrequency ablation in a linear pattern, an irrigated multipolar linear ablation catheter safely delivers contiguous endocardial or epicardial lesions without gaps in a single ablation. This catheter shows promise for decreasing ventricular tachycardia ablation procedure time and improving outcome.
机译:背景技术用于消融瘢痕基室心动过速的常见策略是以线性模式递送多个病变。方法和结果 - 我们测试了一种能够用单个应用产生线性病变的新型线性灌溉多极消融导管的功效。使用新型线性灌溉消融导管进行健康猪接受心外膜和心外膜线性消融;对照动物在带有灌溉的射频导管的3.5厘米的线性图案中接受焦点的焦点。线性导管包含在3.5cm上间隔开的7个灌溉电极,并且可以将^ 25W输送到每个电极。线性消融所需的射线消融明显较低的射频时间而不是焦点烧蚀(56±11与497±110秒; P <0.0001)。在总理理学中,线性(n = 18)心外膜线长于焦点(n = 8)心外膜线(3.3±0.7与2.1±0.9cm; p <0.0005),体积越大(3.8±2.9与1.5±1.6厘米〜3; p = 0.002)。线性(n = 22)和焦点(n = 7)内膜线长度或体积没有区别。在53%的焦线和线性消融线的0%存在间隙(长度为2.8±0.9 mm)。注意到穿孔,蒸汽污染物或血栓。结论 - 与线性图案中的顺序焦点射频消融相比,灌溉的多极线性消融导管安全地提供连续的内膜或心外膜病变,没有间隙在单一消融中。该导管显示了减少心室性心动过速消融程序时间和改善结果的承诺。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号