Variability in split-thickness skin graft depth when using an air-powered dermatome: A paediatric cohort study

机译:使用空气动力皮肤物时分裂厚度皮肤移植深度的变异性:儿科队列研究

获取原文
获取原文并翻译 | 示例
           

摘要

Highlights ? Air dermatomes are not consistent in depth of STSG harvested. ? The median depth is close to the dialed depth at 0.007in. ? Only 50% of grafts are within 0.002in. each side of the dialed depth. ? Pre- or intra-operative variables cannot be identified that may improve consistency. ? This potentially confounds dressing studies using donor site wounds as the test bed. Abstract Aim Split-thickness skin grafts (STSG) taken using calibrated powered dermatomes are assumed to yield a graft of uniform thickness, though this assumption has never been analysed statistically. This study aims to test that assumption in a paediatric population. Method STSGs from a consecutive cohort of paediatric patients were analysed for mean thickness, measured from a central biopsy. All STSGs were taken from the thigh at a dialled thickness of 0.007in. Data were analysed using non-parametric methods. Results There were 140 STSGs taken from 91 children. The median thickness was 6.94 thousandths of an inch, with a spread of thicknesses about this median (IQR 5.05–9.28). There were no significant differences when results were analysed by surgeon, patient age or gender, swipe number within the case, or the number of previous passes with the same blade. Conclusion STSG thickness is inconsistent, with a broad spread about a median value. This study provides no data to suggest there are pre-operative predictors of STSG thickness being significantly more or less than that dialled on a powered dermatome.
机译:强调 ?空气皮肤没有在收获的STSG深度中一致。还中位深度靠近拨号深度0.007英寸。还只有50%的移植物在0.002英寸内。拨号深度的每一侧。还不能识别出或帧内变量,这可以提高一致性。还这可能会使施主部位伤口作为试验台上的敷料研究。摘要采用校准动力性皮肤拍摄的摘要目的厚度皮肤移植物(STSG)以产生均匀厚度的移植物,虽然从未统计地分析了这种假设。本研究旨在测试儿科人群中的假设。从中央活组织检查中测量的分析来自连续的儿科患者患者群体的STSGS。所有STSG都以0.007英寸的拨号厚度取自大腿。使用非参数方法分析数据。结果来自91名儿童中有140个STSGS。中位数厚度为6.94千分之一英寸,厚度散布该中位数(IQR 5.05-9.28)。通过外科医生,患者年龄或性别,在案件内进行刷卡数量的结果,或者以前通过相同刀片的次数的数量,没有显着差异。结论STSG厚度不一致,围绕中位数蔓延。本研究提供无数据建议,STSG厚度的术前预测因子显着或多或少于在动力皮肤上拨入的预测器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号