【24h】

Rapid detection of anti-chromatin autoantibodies in human serum using a portable electrochemical biosensor

机译:使用便携式电化学生物传感器快速检测人血清中的抗染色质自身抗体

获取原文
获取原文并翻译 | 示例

摘要

The current system employed electrochemical sensor technology for the measurement of autoantibodies in human sera. Anti-chromatin antibodies are an early and sensitive indicator of systemic lupus erythematosus (SLE) and are typically quantified using enzyme-linked immunosorbent assays (ELISA). Electrochemical detection compared favorably with ELISA using a sample of 30 SLE sera (r = 0.9), and nonspecific binding by normal serum immunoglobulin was undetectable. The electrochemical sensor assay required <20 min processing time and utilized a hand-held apparatus with a disposable electrode. These results demonstrate the applicability of this technology to the rapid measurement of a clinically relevant analyte with an apparatus of potential simplicity and low cost.
机译:当前的系统采用电化学传感器技术来测量人血清中的自身抗体。抗染色质抗体是系统性红斑狼疮(SLE)的早期敏感指标,通常使用酶联免疫吸附测定(ELISA)进行定量。使用30个SLE血清样品(r = 0.9)进行的电化学检测与ELISA相比具有优势,并且无法检测到正常血清免疫球蛋白的非特异性结合。电化学传感器分析需要少于20分钟的处理时间,并使用了带有一次性电极的手持式设备。这些结果证明该技术可用于具有潜在简单性和低成本的设备快速测量临床相关分析物的适用性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有