Development of a flow standard to enable highly reproducible measurements of deformability of stored red blood cells in a microfluidic device

机译:流动标准的开发能够在微流体装置中能够使储存的红细胞可变形性的高度可再现测量

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND Great deformability allows red blood cells (RBCs) to flow through narrow capillaries in tissues. A number of microfluidic devices with capillary‐like microchannels have been developed to monitor storage‐related impairment of RBC deformability during blood banking operations. This proof‐of‐concept study describes a new method to standardize and improve reproducibility of the RBC deformability measurements using one of these devices. STUDY DESIGN AND METHODS The rate of RBC flow through the microfluidic capillary network of the microvascular analyzer (MVA) device made of polydimethylsiloxane was measured to assess RBC deformability. A suspension of microbeads in a solution of glycerol in phosphate‐buffered saline was developed to be used as an internal flow rate reference alongside RBC samples in the same device. RBC deformability and other in vitro quality markers were assessed weekly in six leukoreduced RBC concentrates (RCCs) dispersed in saline‐adenine‐glucose‐mannitol additive solution and stored over 42?days at 4°C. RESULTS The use of flow reference reduced device‐to‐device measurement variability from 10% to 2%. Repeated‐measure analysis using the generalized estimating equation (GEE) method showed a significant monotonic decrease in relative RBC flow rate with storage from Week 0. By the end of storage, relative RBC flow rate decreased by 22?±?6% on average. CONCLUSIONS The suspension of microbeads was successfully used as a flow reference to increase reproducibility of RBC deformability measurements using the MVA. Deformability results suggest an early and late aging phase for stored RCCs, with significant decreases between successive weeks suggesting a highly sensitive measurement method.
机译:背景技术易变形性允许红细胞(RBC)流过组织中的窄毛细血管。已经开发了许多具有毛细管的微通道的微流体装置,以监测血库操作期间RBC可变形性的与储存有关的损伤。该概念证明研究描述了一种使用这些设备之一标准化和提高RBC可变形测量的再现性的新方法。研究设计和方法通过微血管分析仪(MVA)通过聚二甲基硅氧烷制成的微流体毛细管网络的RBC流量的速率进行测量以评估RBC可变形性。开发磷酸盐缓冲盐水溶液中微珠的悬浮液以用作同一装置中的RBC样品的内部流速参考。每周评估RBC可变形性和其他体外质量标志物,以六种白细胞诱导的RBC浓缩物(RCC)分散在盐水 - 腺嘌呤 - 葡萄糖 - 甘露醇添加剂溶液中并在4℃下储存超过42℃。结果流动参考的使用减少了装置到装置的测量可变性从10%到2%。使用广义估计方程(GEE)方法的重复测量分析显示了相对RBC流速的显着单调减少,从第0周的储存。通过储存结束,相对RBC流速平均降低22?±6%。结论Microbeads的悬浮液被成功用作流动参考,以增加使用MVA的RBC可变形性测量的再现性。可变形结果表明储存RCC的早期和晚期衰老阶段,连续几周之间的显着降低,表明高敏感的测量方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号