Installation of a high-resolution position-sensitive scintillation detector in the small and wide angle neutron scattering instrument (TAIKAN), MLF, J-PARC

机译:在小型角度中子散射仪(Taikan),MLF,J-PARC中安装高分辨率位置敏感闪烁探测器

获取原文
获取原文并翻译 | 示例

摘要

The small and wide angle neutron scattering instrument (TAIKAN) installed at the Material and Life Science Experimental Facility (MLF), Japan Proton Accelerator Research Complex (J-PARC) can cover a Q-range from 5 x 10(-3) to 17 angstrom(-1) simultaneously, where the magnitude of the scattering vector Q is defined by Q = 4 pi sin(theta/2)/lambda (theta and lambda represent the scattering angle and the wavelength, respectively). To extend the accessible low Q-limit (Q(min)) to a value in the order of 10(-4) angstrom(-1), we successfully installed an ultra-small-angle detector with a spatial resolution of approximately 0.41 mm x 0.41 mm in TAIKAN. This detector is a high-resolution position-sensitive detector (HR-PSD), comprising a cross-wired position-sensitive photomultiplier tube and a ZnS/(LiF)-Li-6 scintillator. HR-PSDs of the same type have been installed on steady-state small-angle neutron scattering (SANS) instruments at the research reactor (JRR-3), Japan. Comparing with these previously installed the HR-PSDs, we improved most of the HR-PSD components such as the scintillator, preamplifier, and data acquisition systems to increase its maximum count rate. A maximum count rate of 46 kcps was achieved with 10% counting loss, which is 2.3 times for those installed in the steady-state SANS instruments at JRR-3. Using the ultra-small-angle detector with existing 3He detector banks, TAIKAN can simultaneously cover a Q-range from 8 x 10(-4) angstrom(-1) to 17 angstrom(-1) (over four orders of magnitude in Q).
机译:安装在材料和生命科学实验设施(MLF),日本质子加速器研究络合物(J-PARC)的小和宽角中子散射仪(大观)可以覆盖Q-范围从5×10(-3)至17埃(-1)同时,如果散射矢量Q的幅度由Q = 4 PI罪(希塔/ 2)/λ定义(希塔和λ分别表示散射角和波长)。到可存取低Q-限值(Q(分钟))延伸到一个值中的顺序10(-4)埃(-1),我们成功地安装了一个超小角度检测器具有大约0.41毫米的空间分辨率X0.41毫米在大观。该检测器是一个高分辨率的位置敏感检测器(HR-PSD),包括一个跨有线位置检测型光电倍增管的ZnS /(LIF)-Li-6闪烁体。相同类型的HR-的PSD已安装在处于研究反应器(JRR-3),日本稳态小角中子散射(SANS)仪器。这些以前安装了HR-PSD的比较,我们改进了大部分的HR-PSD分量的,如闪烁器,前置放大器和数据采集系统,以提高其最大计数率。的46个KCPS最大计数速率用10%的计数损失,这是在JRR-3安装在稳态SANS仪器那些2.3倍来实现的。使用超小角度检测器与现有的3He探测器银行,大观可以同时覆盖Q-范围从8×10(-4)埃(-1)到17埃(-1)(在四个Q中数量级)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号