Nowadays, the interest and the number of examination of cerebral angiography and interventional procedure have been significantly increasing. In stent procedure on patients, the important point is to check especially whether the position of stent is complete. Also, the dilution ratio of contrast agent and normal saline (N/S) causes affect the quality of image. Thus, the purpose of this study was to find optimized dilution ratio of contrast agent for acquiring the high quality image with various diameters of self-production vessel model in AAPM head phantom. To evaluate the quality of image, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were used as a factor of quantitative measurement, and Likert scale was used as a factor of qualitative measurement. According to the results, we can acquire excellent result for 1:4 (contrast agent: N/S), 1:5, and 1:5 dilution ratios using 2, 3, and 5 mm vessel diameters, respectively. In addition, quantitative measurement and qualitative measurement came up with same tendency. In conclusion, our results demonstrated that the 1:4 and 1:5 dilution ratio of contrast agent had outstanding results in 2, 3, and 5 mm diameters of vessel model in phantom. (C) 2017 Elsevier GmbH. All rights reserved.
机译:如今,脑血管造影的兴趣和审查次数和介入程序都显着增加。在患者的支架程序中,重要点是检查支架的位置是否完整。而且,造影剂和生理盐水(N / S)的稀释比率会影响图像的质量。因此,本研究的目的是找到造影剂的优化稀释比,以获得具有AAPM Head Phantom的各种自我生产血管模型的高质量图像。为了评估图像的质量,使用信噪比(SNR)和对比度 - 噪声比(CNR)作为定量测量因子,并且李克特量表被用作定性测量因素。根据结果​​,我们可以分别获得1:4(造影剂:N / S),1:5和1:5的稀释比,使用2,3和5mm血管直径获得优异的结果。此外,定量测量和定性测量出现了相同的趋势。总之,我们的结果表明,造影剂的1:4和1:5稀释比在幽灵中的2,3和5mm直径的血管模型中具有出色的结果。 (c)2017年Elsevier GmbH。版权所有。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号