首页> 外文期刊>Current medical research and opinion >Compatibility testing of two types of pen needles with a range of injection pens for diabetes medication.
【24h】

Compatibility testing of two types of pen needles with a range of injection pens for diabetes medication.

机译:两种笔式笔与一系列用于糖尿病药物的注射笔的兼容性测试。

获取原文
获取原文并翻译 | 示例
       

摘要

Compatibility of two types of needles with a variety of durable and prefilled injection pens for diabetes medication was tested by attaching the needles according to ISO 11608-2 and verifying penetration into the cartridge using air shots and two-dimensional X-rays. NovoFine* and NovoFine Autocoverdagger attached correctly to 20 and 19 out of 21 pen types, respectively. Neither needle type attached to Diapen 3.1/3.2, while NovoFine Autocover attached to most, but not all of OptiSet pens.
机译:两种类型的针头与各种用于糖尿病药物的耐用和预填充注射笔的兼容性均通过按照ISO 11608-2安装针头并使用空气喷射和二维X射线验证进入针头盒的方式进行了测试。 NovoFine *和NovoFine Autocoverdagger分别正确连接到21种笔类型中的20和19。 Diapen 3.1 / 3.2均未连接任何针头类型,大多数(但不是全部)OptiSet笔未连接NovoFine Autocover。