...
首页> 外文期刊>Medical Physics >Vertical off‐centering affects organ dose in chest CT CT : Evidence from Monte Carlo simulations in anthropomorphic phantoms
【24h】

Vertical off‐centering affects organ dose in chest CT CT : Evidence from Monte Carlo simulations in anthropomorphic phantoms

机译:垂直偏心会影响胸部CT CT中的器官剂量:来自蒙特卡罗模拟在拟人映像中的证据

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose The objective of our study was to assess the effect of patient vertical off‐centering on organ dose in chest CT with tube current modulation. Methods For this purpose, anthropomorphic phantoms representing adult male, female, and overweight male were scanned on 192‐slice CT scanner at 11 different vertical positions (maximal off‐centering ± 5 cm). Monte Carlo simulations were performed for each of the investigated setup, using tube current values extracted from the raw data, in order to obtain 3D dose distributions. Organ doses were calculated as a function of vertical off‐centering and compared with the reference values, calculated for the phantoms positioned in the gantry isocenter. Image noise was also calculated as a function of phantoms vertical position using few circular regions of interest. Pearson statistical analysis was used to determine the correlation coefficient between image noise and organ dose values with vertical off‐centering. Results Results of our study showed a significant difference in tube currents applied by the CT scanner when the phantom was scanned in off‐centered vertical positions compared to those obtained when the phantom was positioned in the gantry isocenter ( P 0.005). For all investigated phantom configurations the vertical off‐centering below 20 mm in both directions resulted in relative organ dose differences below 7%, while the off‐centering above 40 mm was associated with higher organ dose changes of about 20%. The highest relative dose difference of 38% was observed for the thyroid gland at the lowest table positions. A significant correlation between organ doses for breasts, heart, lungs, thyroid, and liver, and vertical off‐centering (R 2 = 0.909–0.998, P 0.005) was found. The relative dose increase associated with lower table position was more pronounced in peripheral organs: breast and thyroid gland. Image noise behaved opposite to the tube current and organ doses and increased at higher table positions. Conclusion Strong vertical off‐centering in chest CT with tube current modulation results in misoperation of the TCM function affecting both radiation dose and image noise. Therefore, special attention must be paid to a correct patient positioning in order to optimize organ doses and image quality of the respective CT examination.
机译:目的本研究的目的是评估患者垂直偏离围绕器官剂量在胸部CT与管电流调制的效果。方法为了这个目的,拟人化的幻影表示成年男性,女性和男性超重在11个不同的垂直位置进行扫描192切片CT扫描仪(最大偏心±5厘米)。 Monte Carlo模拟中的每个所研究的设置的执行,使用从所述原始数据中提取的管电流值,以便获得3D剂量分布。器官剂量计算为垂直偏心的函数,并且与参考值,定位在台架等中心点幻影计算比较。图像噪声也被计算为使用感兴趣几个圆形区域幻影垂直位置的函数。皮尔逊统计分析来确定与垂直偏心图像噪声和器官的剂量值之间的相关系数。我们的研究结果结果显示在管电流的显著差施加由CT扫描仪,当虚线偏心垂直位置进行扫描时相比,体模被定位在台架等角点获得的那些(P< 0.005)。对于所有研究幻象配置垂直偏心低于20毫米的两个方向上产生了低于7%的相对器官剂量的差异,而偏心以上40毫米用的约20%的更高的器官剂量变化有关。观察到在最低台位置甲状腺38%的最高相对剂量差异。为乳房,心脏,肺,甲状腺,肝和和垂直偏心器官剂量之间的相关性显著(R 2 = 0.909-0.998中,P; 0.005)被发现。具有较低表位置相关联的相对剂量的增加更为明显外周器官:乳腺癌和甲状腺。图像噪声的表现相对的管电流和器官的剂量和在较高的表的位置增加。结论斯特朗垂直偏离定心在胸部CT与既影响辐射剂量和图象噪声的TCM功能的误动作的管电流调制的结果。因此,必须特别注意支付给正确的患者定位,以优化器官剂量和相应的CT检查的图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号