...
首页> 外文期刊>Medical Physics >Investigation of a synthetic diamond detector response in kilovoltage photon beams
【24h】

Investigation of a synthetic diamond detector response in kilovoltage photon beams

机译:千伏电压光子梁合成金刚石探测器响应的研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose An important characteristic of radiation dosimetry detectors is their energy response which consists of absorbed‐dose and intrinsic energy responses. The former can be characterized using Monte Carlo (MC) simulations, whereas the latter (i.e., detector signal per absorbed dose to detector) is extracted from experimental data. Such a characterization is especially relevant when detectors are used in nonrelative measurements at a beam quality that differs from the calibration beam quality. Having in mind the possible application of synthetic diamond detectors (microDiamond PTW 60019, Freiburg, Germany) for nonrelative dosimetry of low‐energy brachytherapy (BT) beams, we determined their intrinsic and absorbed‐dose energy responses in 25–250?kV beams relative to a 60 Co beam, which is usually the reference beam quality for detector calibration in radiotherapy. Material and Methods Three microDiamond detectors and, for comparison, two silicon diodes (PTW 60017) were calibrated in terms of air‐kerma free in air in six x‐ray beam qualities (from 25 to 250?kV) and in terms of absorbed dose to water in a 60 Co beam at the national metrology laboratory in Sweden. The PENELOPE/penEasy MC radiation transport code was used to calculate the absorbed‐dose energy response of the detectors (modeled based on blueprints) relative to air and water depending on calibration conditions. The MC results were used to extract the relative intrinsic energy response of the detectors from the overall energy response. Measurements using an independent setup with a single ophthalmic BEBIG I25.S16 125 I BT seed (effective photon energy of 28?keV) were used as a qualitative check of the extracted intrinsic energy response correction factors. Additionally, the impact of the thickness of the active volume as well as the presence of extra‐cameral components on the absorbed‐dose energy response of a microDiamond detector was studied using MC simulations. Results The relative intrinsic energy response of the microDiamond detectors was higher by a factor of 2 in 25 and 50?kV beams compared to the 60 Co beam. The variation in the relative intrinsic energy response of silicon diodes was within 10% over the investigated photon energy range. The use of relative intrinsic energy response correction factors improved the agreement among the absorbed dose to water values determined using microDiamond detectors and silicon diodes, as well as with the TG‐43 formalism‐based calculations for the 125 I seed. MC study of microDiamond detector design features provided a possible explanation for inter‐detector response variation at low‐energy photon beams by differences in the effective thickness of the active volume. Conclusions MicroDiamond detectors had a non‐negligible variation in the relative intrinsic energy response (factor of 2) which was comparable to that in the absorbed‐dose energy response relative to water at low‐energy photon beams. Silicon diodes, in contrast, had an absorbed‐dose energy dependence on photon energy that varied by a factor of 6, whereas the intrinsic energy dependence on beam quality was within 10%. It is important to decouple these two responses for a full characterization of detector energy response especially when the user and reference beam qualities differ significantly, and MC alone is not enough.
机译:目的,辐射剂量探测器的重要特征是它们的能量响应,由吸收剂量和内在能量反应组成。前者可以使用蒙特卡罗(MC)模拟来表征,而后者(即,每个吸收剂量对检测器的检测器信号)被从实验数据中提取。当在与校准光束质量不同的光束质量下的非相关测量中使用这种表征特别相关。考虑到在低能量近距离放射治疗(BT)梁的非相关剂量测定的非相关剂量测定的合成钻石探测器(Microdiamond PTW 60019,Freiburg,德国)的应用,我们确定了25-250 kV梁相对的内在和吸收剂量响应到一个60 CO梁,通常是放疗中探测器校准的参考光束质量。材料和方法三个微碳化硅探测器和相比,两种硅二极管(PTW 60017)在六个X射线束质量(25至250μl)和吸收剂量方面的空气中校准在瑞典国家计量实验室的60梁中浇水。 PENELOPE / PENEASY MC辐射传输代码用于根据校准条件计算相对于空气和水的检测器的吸收剂量能量响应(基于蓝图建模)。 MC结果用于从整体能量响应中提取检测器的相对内在能量响应。使用与单个眼科Bebig I2525125 I BT种子(28Ω·keV的有效光子能)的测量用作提取的内在能量响应校正因子的定性检查。另外,使用MC仿真研究了主体卷的厚度以及在微水二胺检测器的吸收剂量响应上的存在的影响。结果与60 CO梁相比,微二胺检测器的相对内在能量响应在25和50Ω·kV梁中较高。在研究的光子能量范围内,硅二极管的相对内在能量响应的变化在10%以内。使用相对内在能量响应校正因子的使用改善了吸收剂量的一致性与使用微碳化硅探测器和硅二极管测定的水值,以及125 I种子的TG-43基于TG-43基于形式的计算。 MICODIMOND检测器设计特征的MC研究提供了通过有效厚度的低能量光子束处的检测器响应变化的可能解释。结论微二胺检测器的相对内在能量响应(系数2的因子)具有不可忽略的变化,其与低能量光子梁的吸收剂量响应中的吸收剂量能量响应相当。相反,硅二极管具有对光子能量的吸收剂量能量依赖性,其变化为6,而本质能量对光束质量的依赖性在10%以内。重要的是,特别是当用户和参考光束质量显着不同时,对探测器能量响应的完全表征进行了解耦这两个响应是非常重要的,并且单独使用MC是不够的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号