首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >Direct determination of residual Pluronic F-68 in in-process samples from monoclonal antibody preparations by high performance liquid chromatography
【24h】

Direct determination of residual Pluronic F-68 in in-process samples from monoclonal antibody preparations by high performance liquid chromatography

机译:高效液相色谱法直接测定单克隆抗体制品中的样品中残留的Pluronic F-68

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A simple and sensitive high performance liquid chromatography (HPLC) method was developed to determine residual Pluronic F-68 (PF-68) in in-process samples of monoclonal antibody (MAb) preparations. The method permits the direct injection of proteinaceous samples after simple sample dilution and is able to quantitate as low as 50. mg/L of PF-68 in the presence of up to approximately 30. g/L of protein. The PF-68 molecule was separated on a restricted access reversed phase column using a step gradient and then measured by an evaporative light scattering detector (ELSD). The method was successfully applied to demonstrate PF-68 clearance in MAb purification processes. A modified colorimetric method using liquid-liquid extraction and cobalt thiocyanate to derivatize PF-68 is also described. The results obtained by both the HPLC and colorimetric methods were compared. In addition to its ease of use and simplicity, the HPLC method had better accuracy and higher throughput than the colorimetric method.
机译:建立了一种简单而灵敏的高效液相色谱(HPLC)方法,以测定单克隆抗体(MAb)制备中的样品中残留的Pluronic F-68(PF-68)。该方法允许在简单的样品稀释后直接进样蛋白质样品,并且能够在多达约30 g / L的蛋白质存在下定量低至50 mg / L的PF-68。 PF-68分子使用阶跃梯度在受限进入的反相色谱柱上分离,然后通过蒸发光散射检测器(ELSD)进行测量。该方法已成功应用于证明MAb纯化过程中的PF-68清除率。还描述了一种改进的比色方法,该方法使用液-液萃取和硫氰酸钴将PF-68衍生化。比较了通过HPLC和比色法获得的结果。除了易于使用和简单之外,HPLC法比色度法具有更高的准确性和更高的通量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号