We present a new polarimetric imaging system based on liquid-crystal modulators, a spectrally filtered white-light source, and a CCD camera. The whole Mueller matrix image of the sample is mcasured in approximately 5 s in the transmission mode. The instrument design, together with an original and easy-to-operate calibration procedure, provides high accuracy over a wide spectral range (500-700 nm). This accuracy has been assessed by measurement of a linear polarizer at different orientations and a thick wedged quartz plate as an example of a partially depolarized retarder. Polarimetric images of a stained hepatic biopsy with significant fibrosis have been taken at several wavelengths. The optical properties of Picrosirius Red stained collagen (diattenuation, retardance, and polarizance) have been measured independently from each other between 500 and 700 nm.
机译:我们提出了一种基于液晶调制器,经光谱滤波的白光源和CCD相机的新型偏振成像系统。在传输模式下,将在大约5 s内存储样本的整个Mueller矩阵图像。仪器设计以及原始且易于操作的校准程序,可在宽光谱范围(500-700 nm)内提供高精度。通过测量不同方向的线性偏振器和厚楔形石英板作为部分去偏振的延迟器的示例,可以评估该精度。已在几个波长下拍摄了肝纤维化染色明显的活检的极化图像。 Picrosirius Red染色的胶原的光学特性(衰减,延迟和极化)已在500至700 nm之间相互独立测量。