首页> 外文期刊>Analytical chemistry >Sensitivity Enhancement in the Determination of Volatile Biomarkers in Saliva Using a Mass Spectrometry-Based Electronic Nose with a Programmed Temperature Vaporizer
【24h】

Sensitivity Enhancement in the Determination of Volatile Biomarkers in Saliva Using a Mass Spectrometry-Based Electronic Nose with a Programmed Temperature Vaporizer

机译:基于质谱的电子鼻和程序化温度蒸发器测定唾液中挥发性生物标志物的灵敏度增强

获取原文
获取原文并翻译 | 示例
       

摘要

With a view to improving the sensitivity of direct coupling of a headspace sampler (HS) with a mass spectrometer (MS), here we propose the use of a programmed temperature vaporizer (PTV) in solvent--vent injection mode before the sample is introduced into the MS. This preconcentration scheme has been used for some time in many methods based on gas chromatography (GC), but to the best of our knowledge it has not yet been used in an electronic nose based on MS. The increase in the S/N ratio with the proposed instrumental configuration (HS-PTV/MS) lies between 6.9- and 22-fold. The main advantage of using this injector lies in the fact that it does not involve time-consuming steps. To check the possibilities of this methodology, saliva samples from healthy volunteers and patients with different types of illnesses (including some types of cancer) were analyzed. None of the compounds studied was detected in the samples corresponding to the healthy volunteers. One or more biomarkers, at levels ranging from 13 to 500 μg/L, were found in five of the samples from the patients. Additionally, separative analysis by HS-PTV-GC/MS was performed for confirmatory purposes and both methods provided similar results. The main advantage of the proposed methodology is that no prior chromatographic separation and no sample manipulation are required.
机译:为了提高顶空进样器(HS)与质谱仪(MS)直接耦合的灵敏度,在此我们建议在引入样品之前,在溶剂-进样模式下使用程控温度蒸发器(PTV)进入MS。这种预浓缩方案已在基于气相色谱(GC)的许多方法中使用了一段时间,但据我们所知,它尚未在基于MS的电子鼻中使用。使用建议的仪器配置(HS-PTV / MS),信噪比的增加介于6.9到22倍之间。使用该注射器的主要优点在于它不涉及耗时的步骤。为了检查这种方法的可能性,分析了健康志愿者和患有不同类型疾病(包括某些类型癌症)的患者的唾液样本。在与健康志愿者相对应的样品中未检测到所研究的化合物。在来自患者的五个样本中发现了一种或多种生物标志物,其水平范围为13至500μg/ L。另外,通过HS-PTV-GC / MS进行分离分析是为了验证目的,两种方法都提供了相似的结果。所提出的方法的主要优点是不需要事先进行色谱分离,也不需要样品处理。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号