首页> 外文期刊>Analytical chemistry >Detection of Chemical Weapon Agents and Simulants Using Chemical Ionization Reaction Time-of-Flight Mass Spectrometry
【24h】

Detection of Chemical Weapon Agents and Simulants Using Chemical Ionization Reaction Time-of-Flight Mass Spectrometry

机译:化学电离反应飞行时间质谱检测化学武器剂和模拟物

获取原文
获取原文并翻译 | 示例

摘要

Chemical ionization reaction time-of-flight mass spectrometry(CIR-TOF-MS) has been used for the analysis of prepared mixtures of chemical weapon agents (CWAs) sarin and sulfur mustard. Detection of the CWA simulants 2-chloroethyl ethyl sulfide, triethyl phosphate, and dimethyl methyl phosphonate has also been investigated. Chemical ionization of all the agents and simulants was shown to be possible using the CIR-TOF-MS technique with a variety of reagent ions, and the sensitivity was optimized by variation of instrument parameters. The ionization process was found to be largely unaffected by sample humidity levels, demonstrating the potential suitability of the method to a range of environmental conditions, including the analysis of CWAs in air and in the breath of exposed individuals.
机译:化学电离反应飞行时间质谱(CIR-TOF-MS)已用于分析化学武器剂(CWAs)沙林和芥菜的混合物。还研究了CWA模拟物2-氯乙基乙基硫化物,磷酸三乙酯和膦酸二甲酯甲酯的检测。通过使用CIR-TOF-MS技术和各种试剂离子,可以对所有试剂和模拟物进行化学电离,并且通过改变仪器参数来优化灵敏度。发现电离过程在很大程度上不受样品湿度水平的影响,证明了该方法对一系列环境条件的潜在适用性,包括分析空气中和暴露个体的呼吸中的CWA。