We used thermogravimetric hydrogen reduction and iodometric titration to determine the bulk valence state of cobalt in Co-perovskites containing surface carbonate hydroxide or hydroxyl groups. It could be shown that thermogravimetric hydrogen reduction experiments are very sensitive to volatile surface groups, but due to their volatility, they can be specified and the bulk valence state of cobalt can still be deduced from these experiments. The iodometric titration is less sensitive to small volatile surface impurities, but precaution has to be taken that oxygen or iodine does not escape from the solution during dissolution of the sample. Best results were obtained if the sample was titrated during dissolution in a closed argon floated titration apparatus. We tested the two methods using LaCoO3 perovskite as a sample with a known valence state. Both methods delivered satisfactory results, and the valence state could be determined with an accuracy of better than 1%.
机译:我们使用热重氢还原法和碘量滴定法测定了含有表面碳酸盐氢氧化物或羟基的钙钛矿中钴的整体价态。可以证明,热重氢还原实验对挥发性表面基团非常敏感,但是由于它们的挥发性,可以指定它们,并且仍然可以从这些实验中推导出钴的整体价态。碘量滴定法对较小的挥发性表面杂质较不敏感,但必须采取预防措施,在样品溶解期间,氧气或碘不会从溶液中逸出。如果样品在封闭的氩气浮动滴定仪中溶解时进行滴定,则可获得最佳结果。我们使用钙钛矿型LaCoO3作为具有已知价态的样品测试了这两种方法。两种方法均提供了令人满意的结果,并且价态的准确度可超过1%。