Point-of-care testing (POCT) for blood glucose is widely used in different clinical settings and, thanks to its rapid turnaround time (TAT), it is commonly recommended in patients with critical conditions calling for urgent monitoring. However data in the literature on its reliability and safety are scarce. The aim of the present study was therefore to investigate the type, frequency and clinical risks associated with POCT using portable glucose meters in the hospital setting by evaluating the post-analytical phase when the data are manually transcribed in patients' record. We checked all registrations performed by nurses of 2 selected wards of the hospital during a period of 30 consecutive days, identifying 1966 blood glucose determinations. We found that data obtained were often not reported in patients' files (12.1% of results are missed), the time of blood sampling was recorded in imprecise manner in 7.2% of cases and the glucose concentration was incorrectly reported in 3.2% of results. Although in the post-analytical phase the frequency of incomplete or incorrect data was high, no adverse events were found to be related to this type of error.
机译:血糖即时检测(POCT)在不同的临床环境中得到广泛使用,并且由于其快速周转时间(TAT),通常推荐用于需要紧急监测的危重病患者。但是,关于其可靠性和安全性的文献资料很少。因此,本研究的目的是通过评估在患者记录中手动转录数据后的分析后阶段,来研究在医院环境中使用便携式血糖仪进行POCT的类型,频率和临床风险。我们连续30天检查了医院2个选定病房的护士进行的所有登记,确定了1966年的血糖测定值。我们发现,获得的数据通常没有记录在患者档案中(漏检了12.1%的结果),在7.2%的病例中以不准确的方式记录了采血时间,而在3.2%的结果中错误地报告了葡萄糖浓度。尽管在分析后阶段,数据不完整或不正确的频率很高,但未发现与此类错误相关的不良事件。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号