Continuous improvement of medical test reliability using reference methods and matrix-corrected target values in proficiency testing schemes: application to glucose assay.

机译:在能力验证计划中使用参考方法和经过矩阵校正的目标值来不断提高医学测试的可靠性:应用于葡萄糖测定。

获取原文
获取原文并翻译 | 示例
       

摘要

The reliability of biological tests is a major issue for patient care in terms of public health that involves high economic stakes. Reference methods, as well as regular external quality assessment schemes (EQAS), are needed to monitor the analytical performance of field methods. However, control material commutability is a major concern to assess method accuracy. To overcome material non-commutability, we investigated the possibility of using lyophilized serum samples together with a limited number of frozen serum samples to assign matrix-corrected target values, taking the example of glucose assays. Trueness of the current glucose assays was first measured against a primary reference method by using human frozen sera. Methods using hexokinase and glucose oxidase with spectroreflectometric detection proved very accurate, with bias ranging between -2.2% and +2.3%. Bias of methods using glucose oxidase with spectrophotometric detection was +4.5%. Matrix-related bias of the lyophilized materials was then determined and ranged from +2.5% to -14.4%. Matrix-corrected target values were assigned and used to assess trueness of 22 sub-peer groups. We demonstrated that matrix-corrected target values can be a valuable tool to assess field method accuracy in large scale surveys where commutable materials are not available in sufficient amount with acceptable costs.
机译:就涉及高经济利益的公共卫生而言,生物测试的可靠性是患者护理的主要问题。需要参考方法以及常规的外部质量评估方案(EQAS)来监视现场方法的分析性能。但是,控制材料的可交换性是评估方法准确性的主要问题。为了克服材料的不可交换性,我们以葡萄糖测定为例,研究了将冻干血清样品与有限数量的冷冻血清样品一起使用以指定基质校正目标值的可能性。首先通过使用人冷冻血清对照主要参考方法测量当前葡萄糖测定的真实性。使用己糖激酶和葡萄糖氧化酶结合分光光度法检测的方法非常准确,偏差范围为-2.2%至+ 2.3%。使用葡萄糖氧化酶和分光光度法检测的方法的偏倚为+ 4.5%。然后确定冻干材料的基质相关偏倚,范围为+ 2.5%至-14.4%。分配经过矩阵校正的目标值,并将其用于评估22个子对等组的真实性。我们证明了矩阵校正后的目标值可以作为评估大规模调查中现场方法准确性的有价值的工具,在这些调查中,可兑换材料的数量不足且成本尚可接受。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号