PURPOSE: To determine the ability of the potential acuity meter (PAM) to predict potential visual acuity through lenticular opacities in a group of cataract patients without significant other ocular pathology. METHODS: This prospective study undertook PAM measurements on each eye of 84 subjects during preoperative hospital visits to either A Scan or Primary Care clinics. PAM results were compared to best corrected visual acuity findings obtained at least 6 weeks post-cataract extraction. RESULTS: Comparison of PAM result to visual outcome showed poor correlation (rho = 0.41, P = 0.0005). Using predictive estimates, 49% of subjects' PAM results underestimated visual outcome as determined by a difference of greater than 2 Snellen acuity lines in these results. Factors such as preoperative vision level, cataract type and pupil dilation did not significantly influence the predictive power of the PAM. Preoperative pinhole results were also correlated with visual outcome (rho = 0.42, P = 0.0006). Vison in the fellow eye of subjects with 6/9 or better was correlated to the PAM result indicating a moderate to good predictive power in 'normal' eyes. CONCLUSIONS: PAM results were considered to be poor for a clinical test. The results of this study indicate that this instrument has only a limited usefulness as part of the standard preoperatve examination in patients undergoing cataract extraction.
机译:目的:确定潜在屈光度计(PAM)通过一群白内障患者的双凸浑浊预测潜在视力的能力,而无其他明显的眼部病理。方法:这项前瞻性研究在术前到A Scan或基层医疗诊所就诊的84位受试者的每只眼睛进行了PAM测量。将PAM结果与白内障摘除后至少6周获得的最佳矫正视力检查结果进行比较。结果:PAM结果与视觉结果的比较表明相关性较差(rho = 0.41,P = 0.0005)。使用预测性估计,受试者的PAM结果中有49%的人低估了视觉结果,这是由这些结果中大于2个Snellen敏锐度的差异所决定的。术前视力水平,白内障类型和瞳孔扩大等因素并未显着影响PAM的预测能力。术前针孔结果也与视觉结果相关(rho = 0.42,P = 0.0006)。视力在6/9或更高的受试者的同眼中与PAM结果相关,表明“正常”眼睛的预测能力中等到良好。结论:对于临床测试,PAM结果被认为很差。这项研究的结果表明,该仪器作为接受白内障摘除术的患者术前标准检查的一部分,其用途有限。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号