BACKGROUND: This study was conducted to assess the diurnal variation in ocular hysteresis, as measured by the Ocular Response Analyser to establish a relationship between diurnal hysteresis variation and diurnal intraocular pressure (IOP) variation. METHODS: Forty-two normal eyes of 21 colleagues and staff in a teaching hospital in Birmingham, UK, were recruited. The IOP and hysteresis were measured by the Ocular Response Analyser. The central corneal thickness (CCT) was measured using a hand-held ultrasonic pachymeter in the mid-pupillary axis. RESULTS: The mean ocular hysteresis at 8 am was 12.7 +/- 2.3 mmHg, at 11 am was 12.2 +/- 2.0 mmHg, at 2 PM was 12.7 +/- 2.1 mmHg and at 5 PM was 12.7 +/- 1.7 mmHg; the difference between the values at any time of measurement was not statistically significant (P > 0.9, repeated measures). IOP as measured by non-contact tonometry was 18.4 +/- 2.8 mmHg, 17.9 +/- 3.3 mmHg, 16.9 +/- 3.1 mmHg and 16.8 +/- 3.2 mmHg, respectively, for the same time period; the difference between the values in the morning and afternoon was statistically significant (P < 0.0001, repeated measures). The CCT was 548.8 +/- 29.5 microm, 547.0 +/- 31.4 microm, 548.2 +/- 29.6 microm and 548.6 +/- 29.4 microm, respectively; the difference between the values was not statistically significant at any time points. Multiple regression analysis showed the relationship between IOP and hysteresis was not statistically significant (P = 0.9). CONCLUSION: The ocular hysteresis reading was almost constant throughout the day, whereas the IOP readings showed highest values in the morning with a reducing trend being lowest in the afternoon. The CCT values were almost stable throughout the day. IOP appears to vary independently of a variation in hysteresis or CCT.
机译:背景:本研究旨在评估眼迟滞的昼夜变化,该变化通过眼反应分析仪进行测量,以建立昼夜滞后变化与昼夜眼内压(IOP)变化之间的关系。方法:招募了英国伯明翰一家教学医院的21名同事和工作人员的42只正常眼睛。通过眼响应分析仪测量IOP和滞后。使用手持式超声测厚仪在瞳孔中轴测量中央角膜厚度(CCT)。结果:上午8点的平均眼滞为12.7 +/- 2.3 mmHg,上午11点为12.2 +/- 2.0 mmHg,下午2点为12.7 +/- 2.1 mmHg,下午5点为12.7 +/- 1.7 mmHg。任何时候测量的值之间的差异均无统计学意义(P> 0.9,重复测量)。通过非接触眼压计测得的IOP在同一时期分别为18.4 +/- 2.8 mmHg,17.9 +/- 3.3 mmHg,16.9 +/- 3.1 mmHg和16.8 +/- 3.2 mmHg。上午和下午的值之间的差异具有统计学意义(P <0.0001,重复测量)。 CCT分别为548.8 +/- 29.5微米,547.0 +/- 31.4微米,548.2 +/- 29.6微米和548.6 +/- 29.4微米。在任何时间点,值之间的差异在统计上都不显着。多元回归分析表明,眼压和滞后之间的关系在统计学上不显着(P = 0.9)。结论:全天眼迟滞读数几乎恒定,而IOP读数在早晨显示最高值,而下午下降的趋势则最低。 CCT值全天几乎稳定。 IOP似乎独立于迟滞或CCT的变化而变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号