We evaluated the usefulness of an automated hematology analyzer, SE-9000, to determine the optimal day for harvesting peripheral blood stem cells. Both immature information (IMI) positive cells (IMI+ cells) and hematopoietic progenitor cell (HPC) positive cells (HPC+ cells), which were obtained from the IMI scatter gram of the analyzer, correlated well with CD34 + cells in peripheral blood. Although IMI+ cells sometimes showed false positive data due to toxic granulation of neutrophils and other factors, HPC+ cells apparently did not produce these errors. Thus, the scatter gram should be checked carefully, when IMI+ cells are used. IMI or HPC data obtained immediately before peripheral blood stem cell harvest showed significant correlation with the number of harvested CD34 cells, and could be used to determine the optimal time for harvest.
机译:我们评估了自动血液分析仪SE-9000的有效性,以确定收获外周血干细胞的最佳日期。从分析仪的IMI散射图中获得的未成熟信息(IMI)阳性细胞(IMI +细胞)和造血祖细胞(HPC)阳性细胞(HPC +细胞)与外周血中的CD34 +细胞相关性很好。尽管由于中性粒细胞的毒性颗粒化和其他因素,IMI +细胞有时会显示假阳性数据,但HPC +细胞显然不会产生这些错误。因此,在使用IMI +单元时,应仔细检查散射图。在采集外周血干细胞之前立即获得的IMI或HPC数据显示与采集的CD34细胞数量显着相关,可用于确定最佳采集时间。