首页> 外文期刊>Journal of Wuhan University of Technology. Materials Science edition >Determination of Trace Selenium in Electrolytic Manganese by Graphite Furnace Atomic Absorption Spectrometry
【24h】

Determination of Trace Selenium in Electrolytic Manganese by Graphite Furnace Atomic Absorption Spectrometry

机译:石墨炉原子吸收光谱法测定电解锰中的痕量硒

获取原文
获取原文并翻译 | 示例
       

摘要

The effects of four types of graphite tube and five matrix modifiers on the determination of selenium by graphite furnace atomic absorption spectrometry were compared. The results show that platform thermolysis coat graphite tube and magnesium nitrate and cobaltco as matrix modifer can get a high sensitivity and a good recovery. The optimized working conditions and interference in the determination were invesigated. This result is consistent with that of XRF. The recovery is from 100.8 percent to 102.2 percent ,the relative standard deviation is from 3.47 percent to 5.56 percent (n=9),and the detection limit of selenium is 378 pg (C=44.5 mug/g to 97.3 mug/g.). The proposed method can be applied to the rapid determination of selenium in electrolytic manganese.
机译:比较了四种石墨管和五种基体改性剂对石墨炉原子吸收光谱法测定硒的影响。结果表明,平台热解涂层石墨管和硝酸镁和钴作为基质改性剂可获得较高的灵敏度和良好的回收率。研究了最佳的工作条件和测定中的干扰。此结果与XRF一致。回收率为100.8%至102.2%,相对标准偏差为3.47%至5.56%(n = 9),硒的检出限为378 pg(C = 44.5杯/克至97.3杯/克)。 。该方法可用于电解锰中硒的快速测定。