Accuracy of whole blood glucose measurement when venous catheter blood samples are used on glucose meters.

机译:当在血糖仪上使用静脉导管血液样本时,全血糖测量的准确性。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Previous studies have found positive bias and frequent outliers when central venous catheter (CVC) whole blood is used to dose glucose meters. We designed a study to determine whether positive bias and outliers with CVC whole blood glucose samples are due to exogenous glucose contamination of CVC samples, inherent bias and imprecision of glucose meters, or properties of CVC whole blood that interfere with the function of some glucose meters. METHODS: We studied the relationship between venous whole blood and venous plasma glucose drawn by venipuncture to CVC whole blood and CVC plasma glucose in 50 hospitalized patients. In 27 patients whole blood glucose was measured on both the Accu-Chek Inform (Roche Diagnostics, Indianapolis, IN) and StatStrip (Nova Biomedical, Waltham, MA). RESULTS: By comparing CVC plasma to venous plasma glucose, we determined that contamination of CVC samples with exogenous glucose was uncommon. On the Inform meter outliers were approximately twice as common with CVC whole blood compared to venous whole blood. In 27 patients who had CVC whole blood analyzed by both Inform and StatStrip, outliers occurred approximately twice as often on the Inform compared to the StatStrip. Accounting for CVC samples contaminated with exogenous glucose, outliers on the StatStrip did not occur significantly more often using CVC whole blood compared to venous whole blood. CONCLUSIONS: Properties unique to CVC whole blood differentially affect glucose meter bias and imprecision. Device selection is critical in practices that wish to use CVC whole blood to monitor glucose concentration in hospitalized patients.
机译:背景:以前的研究发现,当使用中央静脉导管(CVC)全血计量血糖仪时,存在正偏见和异常值。我们设计了一项研究,以确定CVC全血糖样品的正偏差和异常值是由于CVC样品的外源葡萄糖污染,血糖仪的固有偏差和不精确性,还是由于CVC全血的特性干扰了某些血糖仪的功能。方法:我们研究了50例住院患者的静脉全血与静脉穿刺抽取静脉血浆葡萄糖与CVC全血和CVC血浆葡萄糖之间的关系。在Accu-Chek通知单(Roche Diagnostics,Indianapolis,IN)和StatStrip(Nova Biomedical,Waltham,MA)上测量了27位患者的全血糖。结果:通过比较CVC血浆和静脉血浆葡萄糖,我们确定外源葡萄糖污染CVC样品的情况并不常见。在Inform仪表上,与静脉全血相比,CVC全血的异常值大约两倍。在通过Inform和StatStrip进行CVC全血分析的27例患者中,与StatStrip相比,在Inform上出现异常值的频率大约是两倍。考虑到被外源葡萄糖污染的CVC样品,与静脉全血相比,使用CVC全血时StatStrip上的异常值发生的可能性并不明显。结论:CVC全血特有的特性差异性地影响血糖仪的偏倚和不准确性。在希望使用CVC全血监测住院患者血糖浓度的实践中,设备选择至关重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号