Reduction in duration of hypoglycemia by automatic suspension of insulin delivery: The in-clinic aspire study

机译:通过自动暂停胰岛素输送来减少低血糖持续时间:这项临床上的向往研究

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Background: The efficacy of automatic suspension of insulin delivery in induced hypoglycemia among subjects with type 1 diabetes was evaluated. Subjects and Methods: In this randomized crossover study, subjects used a sensor-augmented insulin pump system with a low glucose suspend (LGS) feature that automatically stops insulin delivery for 2 h following a sensor glucose (SG) value ≤70 mg/dL. Subjects fasted overnight and exercised until their plasma glucose (measured with the YSI 2300 STAT Plus? glucose and lactate analyzer [YSI Life Sciences, Yellow Springs, OH]) value reached ≤85 mg/dL on different occasions separated by washout periods lasting 3-10 days. Exercise sessions were done with the LGS feature turned on (LGS-On) or with continued insulin delivery regardless of SG value (LGS-Off). The order of LGS-On and LGS-Off sessions was randomly assigned. YSI glucose data were used to compare the duration and severity of hypoglycemia from successful LGS-On and LGS-Off sessions and to estimate the risk of rebound hyperglycemia after pump suspension. Results: Fifty subjects attempted 134 sessions, 98 of which were successful. The mean±SD hypoglycemia duration was less during LGS-On than during LGS-Off sessions (138.5±76.68 vs. 170.7±75.91 min, P=0.006). During LGS-On compared with LGS-Off sessions, mean nadir YSI glucose was higher (59.5±5.72 vs. 57.6±5.69 mg/dL, P=0.015), as was mean end-observation YSI glucose (91.4±41.84 vs. 66.2±13.48 mg/dL, P<0.001). Most (53.2%) end-observation YSI glucose values in LGS-On sessions were in the 70-180 mg/dL range, and none was >250 mg/dL. Conclusions: Automatic suspension of insulin delivery significantly reduced the duration and severity of induced hypoglycemia without causing rebound hyperglycemia.
机译:背景:评估了在1型糖尿病患者中,胰岛素自动悬浮在诱导性低血糖中的功效。受试者和方法:在这项随机交叉研究中,受试者使用具有低葡萄糖悬浮(LGS)功能的传感器增强型胰岛素泵系统,该传感器在传感器葡萄糖(SG)值≤70mg / dL后自动停止2小时的胰岛素输送。受试者禁食过夜并进行运动,直到在不同的情况下其血浆葡萄糖(通过YSI 2300 STAT Plus?葡萄糖和乳酸分析仪[YSI Life Sciences,Yellow Springs,OH]测量)的值达到≤85mg / dL为止,并持续3个洗脱期10天。运动练习是在LGS功能开启(LGS-On)或持续胰岛素输送的情况下进行的,而与SG值无关(LGS-Off)。 LGS-On和LGS-Off会话的顺序是随机分配的。 YSI葡萄糖数据用于比较成功进行LGS-On和LGS-Off疗程后低血糖的持续时间和严重程度,并评估泵悬吊后反弹性高血糖的风险。结果:50名受试者尝试了134次训练,其中98次成功。 LGS-On期间的平均±SD低血糖持续时间少于LGS-Off期间(138.5±76.68 vs. 170.7±75.91 min,P = 0.006)。在LGS-On期间与LGS-Off期间相比,平均天底YSI葡萄糖更高(59.5±5.72 vs. 57.6±5.69 mg / dL,P = 0.015),平均观察到的YSI葡萄糖也较高(91.4±41.84 vs. 66.2)。 ±13.48 mg / dL,P <0.001)。在LGS-On疗程中,大多数(53.2%)的观察结束时YSI葡萄糖值在70-180 mg / dL范围内,没有一个> 250 mg / dL。结论:自动停止胰岛素输送可显着减少诱导性低血糖的持续时间和严重程度,而不会引起反弹性高血糖。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号