首页> 外文期刊>Diabetes technology & therapeutics >Evaluation of a glucose meter in determining the glycemic index of chinese traditional foods.
【24h】

Evaluation of a glucose meter in determining the glycemic index of chinese traditional foods.

机译:在确定中国传统食品的血糖指数中使用血糖仪的评估。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: The purpose of this study was to evaluate whether the Glucometer Elite((R)) meter (Bayer Diagnostics, Kyoto, Japan) (POG) could be used for determining the glycemic index (GI) values of Chinese traditional foods, compared with the YSI (Yellow Springs, OH) glucose analyzer (YSI). METHODS: After consumption of either glucose or one of 11 test foods, the 2-h postprandial glucose responses were measured separately by POG and YSI. GI values and incremental area under the blood glucose response curve (IAUC) values were then calculated to make comparisons between POG and YSI. RESULTS: The mean glucose concentration measured by POG was higher than that measured by YSI (5.54 +/- 0.03 vs. 5.30 +/- 0.03 mmol/L, P < 0.01). The mean IAUC calculated by POG data was also higher than that calculated by YSI data (112.93 +/- 4.04 vs. 106.76 +/- 4.06 mmol.min/L, P < 0.01). However, there were no differences in the mean GI value determined by POG and YSI data for each test food. CONCLUSIONS: Although the glucose concentrations measured by POG were higher than that measured by YSI glucose analyzer, it seemed to be an appropriate instrument to determine the GI values of Chinese traditional foods. However, this conclusion should be cautiously applied to other kinds of glucose meters.
机译:背景:本研究的目的是评估与传统的食品相比,血糖仪(日本京都拜耳诊断公司(POG))是否可用于确定中国传统食品的血糖指数(GI)值。 YSI(Yellow Springs,OH)葡萄糖分析仪(YSI)。方法:食用葡萄糖或11种测试食物之一后,分别通过POG和YSI测量餐后2小时的葡萄糖反应。然后计算GI值和血糖反应曲线(IAUC)值下的增量面积,以比较POG和YSI。结果:POG测得的平均葡萄糖浓度高于YSI测得的平均葡萄糖浓度(5.54 +/- 0.03 vs. 5.30 +/- 0.03 mmol / L,P <0.01)。通过POG数据计算出的IAUC平均值也高于通过YSI数据计算出的平均值(112.93 +/- 4.04对106.76 +/- 4.06 mmol.min / L,P <0.01)。但是,每种测试食品的POG和YSI数据确定的平均GI值没有差异。结论:尽管POG测得的葡萄糖浓度高于YSI血糖分析仪测得的葡萄糖浓度,但它似乎是确定中国传统食品GI值的合适工具。但是,该结论应谨慎应用于其他类型的血糖仪。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号