Background: Large-diameter metal-on-metal articulations reportedly improve stability and wear in THAs. However, some reports suggest some patients have unexplained hip and early failures with these implants. Thus, the potential benefits may be offset by these concerns. However, the incidence of these problems is not clearly established. Questions/purposes: We therefore assessed hip pain, function, osteolysis, and complications in patients with large-diameter metal-on-metal THA. Patients and Methods: We retrospectively reviewed 611 patients who had 681 large-diameter metal-on-metal THAs with the same cup and head design. The average age at operation was 62 years, 53% of the THAs were in men, and the average body mass index was 32 kg/m 2. The diagnosis was osteoarthritis in 92% of the THAs. The minimum followup was 24 months (mean, 37 months; range, 24-60 months). Results: Nine of the 611 patients (1.5%) experienced moderate or severe pain in the hip region that we considered to be coming from an extraarticular source in each case. Harris hip scores for pain averaged 42 points. Total Harris hip scores averaged 93 points. Cup abduction averaged 42°, and cup anteversion averaged 26°. There were no infections. Three cups (0.4%) were considered radiographically loose. All were secondary to inadequate seating of the shell. Conclusion: Our observations suggest with this implant the concerns of higher incidences of groin pain, early failures, and adverse tissue reactions were not confirmed. Early successes or failures with large-diameter metal-on-metal articulations may be implant specific. Level of Evidence: Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
机译:背景:据报道,大直径的金属对金属的连接提高了THA的稳定性和耐磨性。但是,一些报告表明,有些患者无法解释这些植入物的髋关节和早期失败。因此,潜在的利益可能会被这些担忧所抵消。但是,这些问题的发生率尚不清楚。问题/目的:因此,我们评估了大直径金属对金属THA患者的髋部疼痛,功能,骨溶解和并发症。患者和方法:我们回顾性分析了611名患者,他们有681个大直径金属对金属THA,且杯头设计相同。手术的平均年龄为62岁,其中53%的THA为男性,平均体重指数为32 kg / m 2。诊断为92%的THA为骨关节炎。最小随访时间为24个月(平均37个月;范围24-60个月)。结果:611例患者中有9例(1.5%)在髋部区域经历了中度或重度疼痛,我们认为这两种情况均来自关节外。哈里斯的髋关节疼痛评分平均为42分。哈里斯的总髋关节平均得分为93分。杯外展平均为42°,杯前倾平均为26°。没有感染。三杯(0.4%)被认为在射线照相上松动。所有这些都是由于外壳放置不足所致。结论:我们的观察表明,对于这种植入物,腹股沟痛的发生率较高,早期衰竭以及不良的组织反应尚未得到证实。大直径金属对金属铰接的早期成功或失败可能是特定于植入物的。证据级别:IV级,治疗研究。有关证据水平的完整说明,请参见《作者指南》。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号