PURPOSE: To compare the degree of hemolysis generated during use of the ultrasound-based OmniSonics OmniWave Endovascular System and the Possis AngioJet connected to an Xpeedior-6 catheter. MATERIALS AND METHODS: Twelve 64-145-kg Yorkshire pigs were used as the model because their weight, blood volume, and vessel diameters are comparable to that of adult humans. Six pigs were treated with the OmniWave and six with the AngioJet; half of the pigs in each group were treated in the contralateral iliofemoral arteries and half in the ipsilateral iliofemoral veins. Devices were activated for 10 minutes in patent vessels to represent a worst-case scenario for hemolysis generation. Plasma-free hemoglobin (PfHgb) levels, red blood cell counts, white blood cell (WBC) counts, and platelet counts were measured before and after the procedure, and PfHgb levels were measured every 2 minutes during treatment. RESULTS: The OmniWave generated significantly lower PfHgb levels than the AngioJet (average, 228 vs 1,367 mg/dL; P < .001). The location of activation (arterial vs venous) had no significant effect (P = .575). Compared with OmniWave use, AngioJet use was associated with significantly greater changes from pre- to posttreatment values in WBCs (P = .002), platelets (P = .004), and creatinine (P = .002). AngioJet use was associated with a significant postprocedure decrease in the WBC count (mean, -8,000/microl; 95% confidence interval [CI]: -3,830, -12,170; P = .004) and platelet count (mean, -72,000; 95% CI: -25,000, -120,000; P = .011). With the OmniWave device there were no significant differences between pre- and posttreatment WBC count (P = .127), platelet count (P = .493), and creatinine level (P = .317). CONCLUSIONS: The OmniWave generated approximately one-sixth of the hemolysis the AngioJet created. In addition, animals treated with the AngioJet exhibited a decrease in WBC and platelet counts.
机译:目的:比较在使用基于超声波的OmniSonics OmniWave血管内系统和连接到Xpeedior-6导管的Possis AngioJet期间产生的溶血程度。材料与方法:使用十二只64-145公斤约克郡猪作为模型,因为它们的体重,血容量和血管直径与成年猪相当。 OmniWave治疗6头猪,AngioJet治疗6头;每组一半的猪在对侧股动脉中接受治疗,另一半在同侧的of股静脉中接受治疗。设备在专利容器中激活了10分钟,代表发生溶血的最坏情况。在手术前后分别测量无血浆血红蛋白(PfHgb)水平,红细胞计数,白细胞(WBC)计数和血小板计数,并在治疗过程中每2分钟测量一次PfHgb水平。结果:OmniWave产生的PfHgb水平明显低于AngioJet(平均值为228比1,367 mg / dL; P <0.001)。激活的位置(动脉与静脉)无明显影响(P = .575)。与OmniWave的使用相比,AngioJet的使用与白细胞治疗前(P = .002),血小板(P = .004)和肌酐(P = .002)从治疗前到治疗后值的显着增加相关。 AngioJet的使用与手术后WBC计数显着降低(平均值为-8,000 / microl; 95%置信区间[CI]:-3,830,-12,170; P = .004)和血小板计数(平均值为-72,000; 95)有关。 %CI:-25,000,-120,000; P = .011)。使用OmniWave设备,治疗前后的白细胞计数(P = .127),血小板计数(P = .493)和肌酐水平(P = .317)之间没有显着差异。结论:OmniWave大约产生了AngioJet产生的溶血的六分之一。此外,用AngioJet处理的动物的WBC和血小板计数均降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号