Safe magnetic resonance scanning of patients with metallic middle ear implants.

机译:金属中耳植入物患者的安全磁共振扫描。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

OBJECTIVE: To determine the MR scanning risk to patients with otologic implants. DESIGN: We used a repeated-measures study with an additional control measure to assess two aspects of risk; (i) movement of the device in the magnetic field, and (ii) absorption of energy leading to local heating. We used an ex vivo test method that met with international standards. We measured the effects in a Philips Intera Achieva 3 Tesla (T) MR scanner using a Sense Head 8 channel RF coil. SETTING: University-based magnetic resonance research facility. MAIN OUTCOME MEASURES: Heating or displacement of the stapedectomy pistons. RESULTS: No evidence of displacement or heating was found. CONCLUSION: Complying with the ex vivo standard testing protocols, the Schuknecht and McGee wire pistons (device product numbers 140106 and 140108, respectively) were found to be safe in a 3 T MR scanner. These conclusions can be extrapolated to the in vivo case.
机译:目的:确定耳科植入物患者的MR扫描风险。设计:我们使用重复措施研究和其他控制措施来评估风险的两个方面。 (i)装置在磁场中的运动,以及(ii)吸收能量导致局部发热。我们使用了符合国际标准的离体测试方法。我们使用Sense Head 8通道RF线圈在Philips Intera Achieva 3 Tesla(T)MR扫描仪中测量了效果。地点:基于大学的磁共振研究设施。主要观察指标:Heating骨切除术活塞的加热或移位。结果:未发现移位或发热的迹象。结论:符合离体标准测试规程,在3 T MR扫描仪中发现Schuknecht和McGee钢丝活塞(设备产品编号分别为140106和140108)是安全的。这些结论可以外推到体内情况。