Background: Loss of a needle during laparoscopic surgery is a rare but potentially serious adverse event that can cause prolonged operative time and patient harm. Standard recovery techniques currently include instrument count, standard visual search, and plain abdominal x-rays. We developed a laparoscopic instrument to speed the retrieval of lost needles in the abdomen and pelvis. Study Design: We performed in vivo testing of a novel articulating laparoscopic magnet in a porcine model. Three experienced surgeons and 3 inexperienced surgeons conducted 116 needle-retrieval trials with the device and 58 trials with a standard visual approach. Surgeons were blind to the locations of randomly placed surgical needles within the abdominal cavity. Time to recovery was measured and capped at 15 minutes. Analysis was performed using univariate and multivariable methods. Results: The magnetic device was able to retrieve needles significantly faster than the standard approach (2.9 ± 4.0 minutes vs 8.0 ± 6.0 minutes; p < 0.0001). On multivariable analysis, faster recovery time remained independently significant when controlling for surgeon experience, needle size (small, medium, or large), and needle location (by quadrant) (p < 0.0001). There were 2 (2%) injuries to abdominal organs during the device trials and 4 (7%) injuries during the standard trials (p = 0.182). Conclusions: Recovery of lost surgical needles during porcine laparoscopic surgery is safe and feasible with a simple articulating magnetic device. Our initial in vivo experience suggests that recovery is markedly faster using the magnetic device than the standard approach, even in the hands of experienced laparoscopic surgeons. This device will be particularly useful as minimally invasive robotic and single-site surgical techniques are adopted and, in the future, it should be integrated into the standard protocol for locating lost needles during surgery.
机译:背景:腹腔镜手术中失针是一种罕见但潜在的严重不良事件,可导致手术时间延长和患者伤害。目前,标准的恢复技术包括仪器计数,标准的视觉搜索和腹部X线平片。我们开发了一种腹腔镜仪器,以加快腹部和骨盆丢失针的恢复。研究设计:我们在猪模型中进行了新型铰接式腹腔镜磁体的体内测试。 3名经验丰富的外科医生和3名缺乏经验的外科医生使用该设备进行了116次针头取回试验,并使用标准的视觉方法进行了58次试验。外科医生对腹腔内随机放置的手术针的位置视而不见。测量恢复时间并限制在15分钟。使用单变量和多变量方法进行分析。结果:磁性装置能够比标准方法更快地收回针头(2.9±4.0分钟vs 8.0±6.0分钟; p <0.0001)。在多变量分析中,在控制外科医生经验,针头大小(小,中或大)和针头位置(按象限)时,更快的恢复时间仍然独立重要(p <0.0001)。在器械试验中,腹部器官受伤2(2%),在标准试验中,腹部器官受伤4(7%)(p = 0.182)。结论:猪腹腔镜手术中丢失丢失的手术针是安全可行的,只需使用简单的铰接式磁性装置即可。我们最初的体内经验表明,即使在有经验的腹腔镜手术医生的手中,使用磁性装置的恢复也明显比标准方法快。由于采用了微创机器人和单部位手术技术,因此该设备将特别有用,将来,它应集成到用于在手术过程中定位丢失针头的标准协议中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号