Application of Phase Correction to Improve the Interpretation of Crude Oil Spectra Obtained Using 7 T Fourier Transform Ion Cyclotron Resonance Mass Spectrometry

机译:相位校正在改善7 T傅里叶变换离子回旋共振质谱法获得的原油光谱解析度中的应用

获取原文
获取原文并翻译 | 示例
           

摘要

In this study, a phase-correction technique was applied to the study of crude oil spectra obtained using a 7 T Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS). 7 T FT-ICR MS had not been widely used for oil analysis due to the lower resolving power compared with high field FT-ICR MS. For low field instruments, usage of data that has not been phase-corrected results in an inability to resolve critical mass splits of C_3 and SH_4 (3.4 mDa), and ~(13)C and CH (4.5 mDa). This results in incorrect assignments of molecular formulae, and discontinuous double bond equivalents (DBE) and carbon number distributions of S_1, S_2, and hydrocarbon classes are obtained. Application of phase correction to the same data, however, improves the reliability of assignments and produces continuous DBE and carbon number distributions. Therefore, this study clearly demonstrates that phase correction improves data analysis and the reliability of assignments of molecular formulae in crude oil anlayses.
机译:在这项研究中,将相位校正技术应用于使用7 T傅里叶变换离子回旋共振质谱(FT-ICR MS)获得的原油光谱的研究。 7 T FT-ICR MS与高场FT-ICR MS相比具有较低的分辨能力,因此尚未广泛用于油类分析。对于低场仪器,使用未经相位校正的数据将导致无法解析C_3和SH_4(3.4 mDa)以及〜(13)C和CH(4.5 mDa)的临界质量分裂。这导致分子式分配不正确,并且获得了S_1,S_2和烃类的不连续双键当量(DBE)和碳原子数分布。但是,将相位校正应用于相同的数据可提高分配的可靠性,并产生连续的DBE和碳数分布。因此,这项研究清楚地表明,相位校正可以改善数据分析和原油软馏分中分子式分配的可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号