The Planar electrostatic ion trap expands the trapping space of the linear electrostatic ion trap, giving rise to higher tolerance to space charge. A rotational symmetrical design was made, which has a trapping field between two layers of concentric circular electrodes, and the ions are trapped to oscillate around the center plane between the electrodes. The oscillatory motions of the ions were simulated and the field distribution was optimized to achieve isochronous motion against energy spread in R, z, and φ directions. The image charge signal can be picked up by more than one circular electrode and using FFT the mass resolution for the optimized trap can reach 80,000 FWHM. While Fourier transform of the image charge signal generates many high harmonic peaks, the unwanted harmonic peaks can be eliminated by linear combination of image charge signals from multiple pick-up electrodes to give satisfactory results.
机译:平面静电离子阱扩大了线性静电离子阱的捕获空间,从而提高了对空间电荷的耐受性。进行了旋转对称设计,该设计在两层同心圆形电极之间具有捕获场,并且离子被捕获以围绕电极之间的中心平面振荡。模拟了离子的振荡运动,并优化了场分布,以实现针对在R,z和φ方向上传播的能量的等时运动。图像电荷信号可以由多个圆形电极拾取,并且使用FFT,优化阱的质量分辨率可以达到80,000 FWHM。尽管图像电荷信号的傅立叶变换会产生许多高次谐波峰,但可以通过线性组合来自多个拾取电极的图像电荷信号来消除不需要的谐波峰,从而获得令人满意的结果。