An analysis of radiation dose reduction in paediatric interventional cardiology by altering frame rate and use of the anti-scatter grid

机译:通过更改帧频和使用防散射网格减少小儿介入性心脏病的辐射剂量分析

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this work is to investigate removal of the anti-scatter grid and alteration of the frame rate in paediatric interventional cardiology (IC) and assess the impact on radiation dose and image quality. Phantom based experimental studies were performed in a dedicated cardiac catheterisation suite to investigate variations in radiation dose and image quality, with various changes in imaging parameters. Phantom based experimental studies employing these variations in technique identified that radiation dose reductions of 28%-49% can be made to the patient with minimal loss of image quality in smaller sized patients. At present, there is no standard technique for carrying out paediatric IC in the UK or Ireland, resulting in the potential for a wide variation in radiation dose. Dose reductions to patients can be achieved with slight alterations to the imaging equipment with minimal compromise to the image quality. These simple modifications can be easily implemented in clinical practice in IC centres.
机译:这项工作的目的是研究儿科介入性心脏病学(IC)中抗散射网格的去除和帧频的变化,并评估其对辐射剂量和图像质量的影响。在专用的心脏导管套件中进行了基于幻影的实验研究,以研究辐射剂量和图像质量的变化以及成像参数的各种变化。基于幻影的实验研究采用了这些技术上的变化,确定可以对患者进行28%-49%的辐射剂量降低,而较小型患者的图像质量损失却最小。目前,在英国或爱尔兰,尚无用于进行儿科IC的标准技术,导致辐射剂量可能存在很大差异。对患者的剂量减少可以通过对成像设备稍作改动而对图像质量的影响最小。这些简单的修改可以在IC中心的临床实践中轻松实现。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号