PURPOSE: To identify essential considerations for the optimal excimer laser beam for refractive laser surgery and further to determine specific beam parameters fulfilling the essential requirements. METHODS: Starting with the fundamental laser tissue interaction, the various parameters defining the laser beam are analyzed and optimal values for each are evaluated. RESULTS: Gaussian beams generally produce smoother ablated surfaces than "flat top" beams. Optimal ablation efficiency for a Gaussian profile occurs at a peak laser fluence of approximately 450 mJ/cm2. To treat realistic optical aberrations the Gaussian spot size should be < 0.9 mm. The laser repetition rate for such a beam can be multiple hundred hertz without causing thermal damage if a "flying spot" strategy is used. Accurate treatment under such conditions requires both high-fidelity laser spot placement and high-performance eye tracking. CONCLUSIONS: Surgical systems using the characteristics defined in this study should yield excellent treatment outcomes.
机译:目的:确定折光激光手术最佳准分子激光束的基本考虑因素,并进一步确定满足基本要求的特定光束参数。方法:从基本的激光组织相互作用开始,分析定义激光束的各种参数,并评估每种参数的最佳值。结果:高斯光束通常比“平顶”光束产生更光滑的烧蚀表面。高斯轮廓的最佳烧蚀效率在约450 mJ / cm2的峰值激光通量处发生。为了处理真实的光学像差,高斯光斑尺寸应小于0.9 mm。如果使用“飞点”策略,则这种光束的激光重复率可以为数百赫兹,而不会引起热损坏。在这样的条件下进行精确治疗既需要高保真激光点放置又需要高性能的眼动追踪。结论:利用本研究定义的特征的外科手术系统应产生出色的治疗效果。