Coincidence imaging using 2 dual-head gamma-camera systems, with and without attenuation correction.

机译:使用2个双头伽马相机系统进行重合成像,带有和不带有衰减校正。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

OBJECTIVE: Coincidence imaging enhances the potential for imaging a greater number of patients with 18F-FDG in centers that do not have dedicated PET systems. The purpose of this study was to compare, in a clinical setting, coincidence imaging for tumor detection using 2 dual-head gamma-camera systems, one equipped with a 5/8-in. (16 mm) detector (CoDe5) and the other equipped with a newly designed 1-in. (25.4 mm) detector (CoDe8) with an x-ray tube installed in its gantry. METHODS: Thirty consecutive patients were studied by both systems during the same visit and had 4 image sets for comparison: CoDe5 without attenuation correction (CoDe5NC), CoDe8 with (CoDe8AC) and without (CoDe8NC) attenuation correction, and fused coincidence-CT images. The target-to-background ratio (T/Bg ratio) and target-to-nontarget ratio (T/NT ratio) were calculated for each tumor site. RESULTS: On visual assessment, 61 tumor sites were detected on CoDe8AC images. Of these, 59 (97%) were detected on CoDe8NC and 54 (88%) were detected on CoDe5NC images. Fused images improved image interpretation in 10 patients (33%) compared with coincidence images alone. Data added by fusion were of clinical relevance in 6 patients (20%). On quantitative assessment, the number of accepted events by the CoDe8 was significantly higher than that by CoDe5 (5.21 +/- 1.46 million vs. 1.27 +/- 0.36 million, P <0.001). When comparing CoDe5 with CoDe8 images without attenuation correction, the T/Bg and T/NT ratios were significantly higher on the CoDe8 images (P <0.0005 and P <0.0005, respectively). When comparing CoDe8 images with and without attenuation correction, the T/Bg ratio was better on the attenuation-corrected images (P <0.0005). CONCLUSION: Coincidence imaging with 1-in. detectors and attenuation correction improve image quality and, to a lesser extent, the tumor detection rate compared with the 5/8-in. detectors and noncorrected images. The data added by fusion of coincidence images to CT findings were clinically relevant in 20% of the patients.
机译:目的:重合成像增强了在没有专用PET系统的中心对更多18F-FDG患者进行成像的潜力。这项研究的目的是在临床环境中比较使用2个双头伽马相机系统(其中一个配备5/8英寸)进行肿瘤检测的重合成像。 (16毫米)检测器(CoDe5),另一个则配备了新设计的1英寸。 (25.4 mm)检测器(CoDe8),其机架中装有X射线管。方法:在同一次就诊期间,两个系统均对30例连续患者进行了研究,并具有4个图像集进行比较:不带衰减校正的CoDe5(CoDe5NC),带(CoDe8AC)和不带(CoDe8NC)衰减校正的CoDe8和融合的CT重合图像。计算每个肿瘤部位的靶与背景之比(T / Bg比)和靶与非靶之比(T / NT比)。结果:通过视觉评估,在CoDe8AC图像上检测到61个肿瘤部位。其中,在CoDe8NC上检测到59(97%),在CoDe5NC图像上检测到54(88%)。与仅符合图像相比,融合图像改善了10位患者(33%)的图像解释。融合添加的数据在6例患者中具有临床意义(20%)。在定量评估中,CoDe8接受的事件数显着高于CoDe5(5.21 +/- 1.46百万对1.27 +/- 36万),P <0.001)。将CoDe5图像与未进行衰减校正的CoDe8图像进行比较时,CoDe8图像上的T / Bg和T / NT比明显更高(分别为P <0.0005和P <0.0005)。比较具有和不具有衰减校正的CoDe8图像时,在衰减校正后的图像上,T / Bg比更好(P <0.0005)。结论:1英寸重合成像。与5/8英寸相比,检测器和衰减校正可改善图像质量,并在较小程度上改善肿瘤检测率。检测器和未校正的图像。在20%的患者中,通过将符合图像融合到CT发现中而增加的数据在临床上具有相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号