A multicommuted flow system for the determination of dextrose in parenteral and hemodialysis concentrate solutions.

机译:用于测定肠胃外和血液透析浓缩液中葡萄糖含量的多通流系统。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A new method is presented for the automation of the determination of dextrose in parenteral and hemodialysis solutions. The method is based on multicommutation flow analysis (MCFA) and exploits enzymatic reactions providing a colored derivative that is detected spectrophotometrically (Trinder's method). The reagent, comprising glucose oxidase, peroxidase, 4-hydroxybenzoate and a buffer was obtained from a commercial kit for the determination of glycemia. The flow system used three 3-way solenoid valves operating under computer control. The necessary software was developed for the purpose and compiled in QuickBASIC 4.0 The influence of some operating variables (segment size, number of segments and reactor length) was studied. Calibration curves in the range 0-1g/L presented a slight curvature and were fitted with a second-degree polynomial (h=-0.0632C(2)+0.6039C+0.166, r(2)=0.9973, h being the peak-height (absorbance) and C the concentration in g/L). The method was validated by analyzing artificial samples presenting accurately known concentrations of dextrose, and comparing the results with the known value and with value obtained by polarimetry. Recoveries were in the range 96.6-100.2%, and the difference with the polarimetric analysis was in the range 0.1-3.3%. Precision (R.S.D.,%) was better than 2.4%. Sampling frequency of the system was 90 samples/h, with a reagent consumption of 0.14mL per sample.
机译:提出了一种自动测定肠胃外和血液透析溶液中葡萄糖的新方法。该方法基于多换向流分析(MCFA),并利用酶促反应提供了一种有色衍生物,可通过分光光度法对其进行检测(Trinder方法)。该试剂包括葡萄糖氧化酶,过氧化物酶,4-羟基苯甲酸酯和缓冲液,可从商业试剂盒中获得,用于测定血糖。流量系统使用了三个在计算机控制下运行的三通电磁阀。为此目的开发了必要的软件,并在QuickBASIC 4.0中进行了编译。研究了一些操作变量(段大小,段数和反应器长度)的影响。 0-1g / L范围内的校准曲线略有弯曲,并拟合了二次多项式(h = -0.0632C(2)+ 0.6039C + 0.166,r(2)= 0.9973,h为峰值-高度(吸光度)和C浓度,以g / L为单位)。通过分析呈现准确已知葡萄糖浓度的人工样品,并将结果与​​已知值和通过旋光法获得的值进行比较,验证了该方法的有效性。回收率在96.6-100.2%范围内,与偏振分析的差异在0.1-3.3%范围内。精度(R.S.D。,%)优于2.4%。系统的采样频率为90个样品/小时,每个样品的试剂消耗为0.14mL。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号