首页> 美国卫生研究院文献>The Journal of Automatic Chemistry >Multiparametric Flow System for the Automated Determination of Sodium Potassium Calcium and Magnesium in Large-Volume Parenteral Solutions and Concentrated Hemodialysis Solutions
【2h】

Multiparametric Flow System for the Automated Determination of Sodium Potassium Calcium and Magnesium in Large-Volume Parenteral Solutions and Concentrated Hemodialysis Solutions

机译:用于自动测定大剂量肠胃外溶液和浓缩血液透析溶液中钠钾钙和镁的多参数流系统

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A multiparametric flow system based on multicommutation and binary sampling has been designed for the automated determination of sodium, potassium, calcium, and magnesium in large-volume parenteral solutions and hemodialysis concentrated solutions. The goal was to obtain a computer-controlled system capable of determining the four metals without extensive modifications. The system involved the use of five solenoid valves under software control, allowing the establishment of the appropriate flow conditions for each analyte, that is, sample size, dilution, reagent addition, and so forth. Detection was carried out by either flame atomic emission spectrometry (sodium, potassium) or flame atomic absorption spectrometry (calcium, magnesium). The influence of several operating parameters was studied. Validation was carried out by analyzing artificial samples. Figures of merit obtained include linearity, accuracy, precision, and sampling frequency. Linearity was satisfactory: sodium, r2  0.999 (0.5–3.5 g/L), potassium, r2  0.996 (50–150 mg/L), calcium, r2  0.999 (30–120 mg/L), and magnesium, r2  0.999 (20–40 mg/L). Precision (sr, %, n = 5) was better than 2.1%, and accuracy (evaluated through recovery assays) was in the range of 99.8%–101.0% (sodium), 100.8–102.5% (potassium), 97.3%–101.3% (calcium), and 97.1%–99.8% (magnesium). Sampling frequencies (h−1) were 70 (sodium), 75 (potassium), 70 (calcium), and 58 (magnesium). According to the results obtained, the use of an automated multiparametric system based on multicommutation offers several advantages for the quality control of large-volume parenteral solutions and hemodialysis concentrated solutions.
机译:设计了基于多换向和二元采样的多参数流动系统,用于自动测定大剂量肠胃外溶液和血液透析浓缩溶液中的钠,钾,钙和镁。目的是获得一种无需大量修改即可确定四种金属的计算机控制系统。该系统涉及在软件控制下使用五个电磁阀,从而允许为每种分析物建立合适的流动条件,即样品大小,稀释度,试剂添加等。通过火焰原子发射光谱法(钠,钾)或火焰原子吸收光谱法(钙,镁)进行检测。研究了几个操作参数的影响。通过分析人工样品进行验证。获得的品质因数包括线性,精度,精度和采样频率。线性令人满意:钠,r 2 2 2 2 -1 )为70(钠),75(钾),70(钙)和58(镁)。根据获得的结果,基于多换向的自动化多参数系统的使用为大剂量肠胃外溶液和血液透析浓缩溶液的质量控制提供了多个优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号