Validation of a Computer-Assisted Method for Measurement of Radiographic Wear in Total Hip Arthroplasty Using All Polyethylene Cemented Acetabular Components

机译:使用所有聚乙烯骨水泥髋臼组件的全髋关节置换术中射线照相磨损测量的计算机辅助方法的验证

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Although cemented all polyethylene (PE) cups have been routinely used in total hip arthroplasty for decades, no computer-assisted method for measurement of radiographic wear has ever been specifically validated for these implants. Using a validated hip phantom model, AP plain hip radiographs were obtained consecutively for eight simulated wear positions. A version of Martell's Hip Analysis Suite software dedicated to all polyethylene sockets was used by three different examiners of varied experience. Bias (mean, standard deviation and 95% confidence interval limit), repeatability (standard deviation and 95% limit) and reproducibility (standard deviation and 95% limit) for two-dimensional wear measurements were assessed, as recommended by the current ASTM guidelines. Using this protocol, the dedicated software showed an overall mean bias of 0.089 +/- 0.060mm (mean +/- SD), and 0.118mm for 95% CI limit. Repeatability (intra examiner) standard deviation and 95% limit were respectively 0.106mm and 0.292mm. Reproducibility (inter examiner) standard deviation and 95% limit were respectively 0.112mm and 0.308mm. Martell Hip Analysis for all PE cemented cups is a reliable and low-cost instrument in the assessment of wear, despite being less precise than its original version dedicated to cementless components. (c) 2015 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 33:???-???, 2015.
机译:尽管胶合所有聚乙烯(PE)杯已在全髋关节置换术中常规使用了数十年,但尚未针对这些植入物专门验证过计算机辅助的射线照相磨损测量方法。使用经过验证的髋部模型,连续八个模拟佩戴位置获得了AP普通髋部X线照片。三位经验丰富的审查员使用了专门用于所有聚乙烯插座的Martell Hip Analysis Suite软件版本。根据当前ASTM指南的建议,评估了二维磨损测量的偏差(平均值,标准偏差和95%置信区间极限),重复性(标准偏差和95%极限)和再现性(标准偏差和95%极限)。使用此协议,专用软件显示总体平均偏差为0.089 +/- 0.060mm(平均+/- SD),对于95%CI限制为0.118mm。重复性(内部检验者)标准偏差和95%极限分别为0.106mm和0.292mm。再现性(内部审查员)标准偏差和95%限制分别为0.112mm和0.308mm。尽管其精确度低于专用于非水泥组件的原始版本,但适用于所有PE水泥杯的Martell Hip Analysis是一种可靠且低成本的评估磨损的仪器。 (c)2015骨科研究学会。由Wiley Periodicals,Inc.出版。J Orthop Res 33:???-???,2015年。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号