首页> 外文期刊>Journal of orthopaedic science : >The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis.
【24h】

The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis.

机译:头部直径和壁厚对金属髋臼压配合杯和UHMWPE内衬变形的影响:有限元分析。

获取原文
获取原文并翻译 | 示例
           

摘要

To increase the range of motion of total hip endoprostheses, prosthetic heads need to be enlarged, which implies that the cup and/or liner thickness must decrease. This may have negative effects on the wear rate, because the acetabular cups and liners could deform during press-fit implantation and hip joint loading. We compared the metal cup and polyethylene liner deformations that occurred when different wall thicknesses were used in order to evaluate the resulting changes in the clearance of the articulating region.A parametric finite element model utilized three cup and liner wall thicknesses to analyze cup and liner deformations after press-fit implantation into the pelvic bone. The resultant hip joint force during heel strike was applied while the femur was fixed, accounting for physiological muscle forces. The deformation behavior of the liner under joint loading was therefore assessed as a function of the head diameter and the resulting clearance.Press-fit implantation showed diametral cup deformations of 0.096, 0.034, and 0.014?mm for cup wall thicknesses of 3, 5, and 7?mm, respectively. The largest deformations (average 0.084?±?0.003?mm) of liners with thicknesses of 4, 6, and 8?mm occurred with the smallest cup wall thickness (3?mm). The smallest liner deformation (0.011?mm) was obtained with largest cup and liner wall thicknesses. Under joint loading, liner deformations in thin-walled acetabular cups (3?mm) reduced the initial clearance by about 50?%.Acetabular press-fit cups with wall thicknesses of ≤5?mm should only be used in combination with polyethylene liners >6?mm thick in order to minimize the reduction in clearance.
机译:为了增加整个髋关节假体的运动范围,需要扩大假体头部,这意味着必须减小杯和/或衬垫的厚度。这可能对磨损率有负面影响,因为髋臼杯和内衬在压入配合植入和髋关节负荷期间可能变形。我们比较了使用不同壁厚时发生的金属杯和聚乙烯衬里变形,以评估关节区域间隙的变化情况。参数有限元模型利用三种杯和衬里壁厚来分析杯和衬里变形压入式植入骨盆骨后。在固定股骨的同时施加足跟撞击期间产生的髋关节力,这是生理肌肉力的原因。因此,评估了在联合载荷下衬套的变形行为与头部直径和所产生的间隙的关系。对于杯壁厚度为3、5、5的压配合植入,其径向杯变形为0.096、0.034和0.014?mm。和7?mm。厚度为4、6和8μmm的衬管最大变形(平均0.084±0.003μmm)发生在最小的杯壁厚度(3μmm)处。杯形和衬里壁厚最大时,衬里变形最小(0.011?mm)。在联合载荷下,薄壁髋臼杯(3?mm)的内衬变形使初始间隙降低了约50?%。壁厚≤5?mm的髋臼压配合杯只能与聚乙烯内衬结合使用>厚度为6?mm,以最小化间隙的减小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号