We describe a head-stage, with precision microtranslators for the chronic placement of micro-wire electrodes in the neocortex, that minimizes compressive damage to the brain. The head-stage has a diameter of 5.8 mm and allows six electrodes, separated by 450 microm on a hexagonal grid, to be individually and continuously positioned throughout a depth of approximately 3 mm. Suction is used to transiently support the dura against a curved array of tubes that guide and stabilize the electrodes as a means to prevent compression of the neocortex as the electrodes breach the dura. With this headstage we recorded extracellular signals in a rat immediately after surgery. Single-unit waveforms at a given electrode position were stable for at least several hours in the freely behaving animal and were obtained throughout the depth of the neocortex for at least 2 months. Electrophysiological records and histological examination showed that the upper layers of the neocortex were intact and minimally damaged after the implantation.
机译:我们描述了一个具有精密微翻译器的头级设备,用于将微丝电极长期放置在新皮层中,从而最大程度地减少了对大脑的压缩损伤。头台的直径为5.8 mm,并允许六个电极(在六角形网格上相隔450微米)在大约3 mm的深度内单独连续放置。抽吸用于将硬脑膜暂时支撑在弯曲的管阵列上,该管弯曲并引导电极并稳定电极,以防止在电极突破硬脑膜时压缩新皮质。在这个阶段,我们在手术后立即在大鼠中记录了细胞外信号。在自由行为的动物中,给定电极位置的单单位波形在至少几个小时内稳定,并在整个新皮层深度至少2个月内获得。电生理记录和组织学检查显示,植入后新皮层的上层完好无损且受到的损害最小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号