In this paper we present the design and fabrication of the two fundamental components of a fluid handling system intended to be the core of a chemical microanalyzer for measuring heavy metals in ground water. The first fluidic component is a micropump with built-in passive valves. The pump body is fabricated by thermoplastic moulding, whereas the passive valves are obtained by reaction injection moulding. The second fluidic component is an active valve, also fabricated by thermoplastic moulding. Both devices are driven by miniaturized electromagnetic actuators. Experiments demonstrate that the head pressure produced by the pump is 5.5 x 10(3) Pa and the maximum flow rate is 13 ml s(-1). The active valve is designed to be normally closed till an external pressure of 6.5 x 10(3) Pa. [References: 2]
机译:在本文中,我们介绍了流体处理系统的两个基本组件的设计和制造,该系统旨在作为用于测量地下水中重金属的化学微分析仪的核心。第一个流体组件是带有内置被动阀的微型泵。泵体通过热塑成型制造,而被动阀通过反应注射成型获得。第二流体部件是主动阀,该主动阀也通过热塑性模制制造。两种设备均由小型电磁致动器驱动。实验表明,泵产生的扬程压力为5.5 x 10(3)Pa,最大流量为13 ml s(-1)。主动阀设计为常闭,直到外部压力为6.5 x 10(3)Pa。[参考:2]