OBJECTIVES: To monitor the development of the stability of Straumann tissue-level implants during the early phases of healing by resonance frequency analysis (RFA) and to determine the influence of implant surface modification and diameter. MATERIAL AND METHODS: A total of twenty-five 10 mm length implants including 12 SLA RN v4.1 mm implants, eight SLActive RN v4.1 mm implants and five SLA WN v4.8 mm implants were placed. Implant stability quotient (ISQ) values were determined with Osstell mentor at baseline, 4 days, 1, 2, 3, 4, 6, 8 and 12 weeks post-surgery. ISQ values were compared between implant types using unpaired t-tests and longitudinally within implant types using paired t-tests. RESULTS: During healing, ISQ decreased by 3-4 values after installation and reached the lowest values at 3 weeks. Following this, the ISQ values increased steadily for all implants and up to 12 weeks. No significant differences were noted over time. The longitudinal changes in the ISQ values showed the same patterns for SLA implants, SLActive implants and WB implants. At placement, the mean ISQ values were 72.6, 75.7 and 74.4, respectively. The mean lowest ISQ values, recorded at 3 weeks, were 69.9, 71.4 and 69.8, respectively. At 12 weeks, the mean ISQ values were 76.5, 78.8 and 77.8, respectively. The mean ISQ values at all observation periods did not differ significantly among the various types. Single ISQ values ranged from 55 to 84 during the entire healing period. Pocket probing depths of the implants ranged from 1 to 3 mm and bleeding on probing from 0 to 2 sites/implant post-surgically. CONCLUSIONS: All ISQ values indicated the stability of Straumann implants over a 12-week healing period. All implants showed a slight decrease after installation, with the lowest ISQ values being reached at 3 weeks. ISQ values were restored 8 weeks post-surgically. It is recommended to monitor implant stability by RFA at 3 and 8 weeks post-surgically. However, neither implant surface modifications (SLActive) nor implant diameter were revealed by RFA.
机译:目的:通过共振频率分析(RFA)监测Straumann组织级植入物在愈合早期的稳定性,并确定植入物表面修饰和直径的影响。材料和方法:总共放置了25个10 mm长的植入物,包括12个SLA RN v4.1 mm植入物,八个SLActive RN v4.1 mm植入物和五个SLA WN v4.8 mm植入物。在手术后第4天,1、2、3、4、6、8和12周,在基线时用Osstell导师测定植入物稳定性商(ISQ)值。使用未配对的t检验比较植入物类型之间的ISQ值,并使用配对的t检验纵向比较植入物类型内的ISQ值。结果:在康复期间,ISQ在安装后下降了3-4个值,并在3周时达到最低值。此后,所有植入物的ISQ值稳定增加,长达12周。随着时间的推移,没有发现明显的差异。对于SLA植入物,SLActive植入物和WB植入物,ISQ值的纵向变化显示出相同的模式。放置时,平均ISQ值分别为72.6、75.7和74.4。 3周记录的平均最低ISQ值分别为69.9、71.4和69.8。在12周时,平均ISQ值分别为76.5、78.8和77.8。所有观察期的平均ISQ值在各种类型之间没有显着差异。在整个康复期间,单个ISQ值的范围从55到84。植入物的口袋探测深度范围为1到3 mm,手术后探测/植入物的探测范围为0到2个部位。结论:所有ISQ值均表明Straumann植入物在12周愈合期内具有稳定性。所有植入物在安装后均显示出轻微的下降,在3周时达到最低的ISQ值。术后8周恢复ISQ值。建议在术后3周和8周通过RFA监测植入物的稳定性。但是,RFA并未显示出植入物表面修饰(SLActive)或植入物直径。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号