MR imaging of the female pelvis at 3 Tesla: Evaluation of image homogeneity using different dielectric pads.

机译:3特斯拉下女性骨盆的MR成像:使用不同的介电垫评估图像均匀性。

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: To evaluate improvements in image homogeneity in pelvic MR imaging at 3 Tesla (T) using two different dielectric pads. MATERIALS AND METHODS: A total of eight healthy females were scanned using a 3T MR scanner equipped with a body-array coil. Axial and sagittal fast spin-echo T2-weighted images (T2WI) (TR/TE = 3200 msec/94 msec), axial fast spin-echo T1-weighted images (T1WI) (TR/TE = 700 msec/11 msec), and sagittal half-Fourier acquisition single-shot turbo spin-echo (HASTE) images (TR/TE = 3000 msec/100 msec) were performed for pelvic imaging. Sequences were repeated with dielectric pads (consisting of either ultrasound [US] gel or water), and without pads. Three or four regions of interest (ROIs) were placed on fatty tissues and the ratio of minimum to maximum signal intensity (RSI) was calculated as a marker of image homogeneity. RESULTS: RSI was significantly higher on T2WI and T1WI when using dielectric pads than when no pad was used. A similar tendency was observed in RSI on HASTE. No significant difference was found between images with US gel pads and those with water pads. CONCLUSION: Dielectric pads consisting of either US gel or water are effective in improving image homogeneity of the pelvis on 3T MRI. J. Magn. Reson. Imaging 2007. (c) 2007 Wiley-Liss, Inc.
机译:目的:评估使用两个不同的电介质垫在3 Tesla(T)下骨盆MR成像中图像均匀性的改善。材料与方法:使用配备了人体线圈的3T MR扫描仪对八名健康女性进行了扫描。轴向和矢状快速自旋回波T2加权图像(T2WI)(TR / TE = 3200毫秒/ 94毫秒),轴向快速自旋回波T1加权图像(T1WI)(TR / TE = 700毫秒/ 11毫秒),并进行矢状半傅里叶采集单发涡轮自旋回波(HASTE)图像(TR / TE = 3000毫秒/ 100毫秒)进行骨盆成像。使用电介质垫(由超声[US]凝胶或水组成)和不使用电垫重复序列。将三个或四个感兴趣区域(ROI)放置在脂肪组织上,并计算最小信号强度与最大信号强度的比率(RSI)作为图像均匀性的标志。结果:使用介电焊盘时,T2WI和T1WI的RSI明显高于未使用焊盘时的RSI。在HASTE上的RSI中观察到了类似的趋势。使用US胶垫的图像和使用水垫的图像之间没有发现显着差异。结论:由US凝胶或水组成的介电垫可有效改善3T MRI上骨盆的图像均匀性。 J.Magn。雷森影像杂志2007。(c)2007 Wiley-Liss,Inc.。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号