摘要

PURPOSE: To examine the influence of the truncated sampling of k-space data on the accuracy of phase-contrast (PC) flow quantifications in the presence of nearby vessels. MATERIALS AND METHODS: Computer simulations were performed along with some experimental validations on a flow phantom and a normal subject. RESULTS: The accuracy of the PC flow quantification decreased when a second vessel was positioned closely to the examined vessel. The effect strongly depended on the peak flow velocity in the second vessel relative to the velocity encoding (venc) level of the MRI acquisition, and on the position and area of the second vessel. CONCLUSION: Due to the truncated sampling of the k-space data, signal leaked from nearby vessels and distorted the PC flow quantification in the examined vessel. The presented results suggest specific experimental conditions to minimize this flow quantification error.
机译:目的:研究在附近血管存在的情况下,k空间数据的截断采样对相衬(PC)流定量分析的准确性的影响。材料与方法:对流模型和正常对象进行了计算机模拟以及一些实验验证。结果:当第二个血管紧靠检查血管放置时,PC流量定量的准确性降低。效果很大程度上取决于第二血管中相对于MRI采集的速度编码(venc)水平的峰值流速,以及第二血管的位置和面积。结论:由于k空间数据的截断采样,信号从附近的血管泄漏,并使被检测血管中的PC流量定量失真。提出的结果提出了特定的实验条件,以最大程度地减少这种流量定量误差。