Concurrent validity of flexicurve instrument measurements: sagittal skin contour of the cervical spine compared with lateral cervical radiographic measurements.

机译:屈曲仪测量的同时有效性:颈椎的矢状皮肤轮廓与颈椎侧位X线照相测量的比较。

获取原文
获取原文并翻译 | 示例
       

摘要

OBJECTIVES: The aim of this study was to compare flexicurve surface contour measurements of the cervical spine with radiographic measurements of cervical lordosis. METHODS: One examiner evaluated 96 patients with chronic neck pain in neutral posture using a flexible ruler, flexicurve, to measure sagittal contour of the skin over the cervical spine from the external occipital protuberance to the vertebra prominens. The flexicurve skin contour and neutral lateral radiographs were digitized and compared. The flexicurve and radiographs were categorized into height-length ratio, curve angle, curve depth, sum of depths, modified Ishihara's index, and inverse of radius. Mean values, SDs, mean differences, and limits of agreement were calculated. The differences between flexicurve measurement mean values and x-ray mean values were deemed significant if the lower limit of agreement exceeded 15% of the mean values for the x-ray measurements. RESULTS: For all variables, except the height-length ratio, the mean values of the flexicurve variables differed significantly from the corresponding mean values of the radiographic measurements. All Pearson correlation coefficients were in the very poor range (r < 0.15). CONCLUSION: The flexicurve sagittal skin contour measurement has poor concurrent validity compared with established radiographic measurements of the cervical lordosis. The flexicurve tracings always predicted lordosis, overestimated the lordosis compared with x-ray values, and cannot discriminate between radiographic lordosis, straightened, S curves, and kyphotic alignments of the cervical curve.
机译:目的:本研究的目的是比较颈椎弯曲弯曲的表面轮廓测量值与颈椎前凸的影像学测量值。方法:一位检查员使用挠性尺屈曲评估了96例中性姿势的慢性颈痛患者,以测量从外枕骨突起到椎骨隆起的颈椎皮肤矢状轮廓。数字化并比较了弯曲的皮肤轮廓和中性的侧位X线照片。弯曲曲线和射线照片分为高长比,弯曲角度,弯曲深度,深度总和,修正的Ishihara指数和半径的倒数。计算平均值,标准差,均差和一致限。如果一致性的下限超过X射线测量平均值的15%,则弯曲测量平均值和X射线平均值之间的差异被认为是显着的。结果:对于所有变量,除了高长比之外,弯曲曲线变量的平均值与射线照相测量的相应平均值有显着差异。所有Pearson相关系数都在非常差的范围内(r <0.15)。结论:与既定的颈椎前凸影像学检查相比,弯曲的弧矢状皮肤轮廓测量的并发有效性较差。弯曲曲线描记总是预测脊柱前凸,与X射线值相比高估了脊柱前凸,并且无法区分放射线形脊柱前凸,矫直,S曲线和颈椎弯曲的后凸。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号