Validation of Optical Coherence Tomography against Micro-computed Tomography for Evaluation of Remaining Coronal Dentin Thickness

机译:光学相干断层扫描对微计算机断层扫描的验证,以评估剩余冠状牙本质厚度

获取原文
获取原文并翻译 | 示例

摘要

Introduction: Optical coherence tomography (OCT) is a noninvasive modality to obtain in-depth images of biological structures. A dental OCT system has become available for chairside application. This in vitro study hypothesized that swept-source OCT can be used to measure the remaining dentin thickness (RDT) at the roof of the dental pulp chamber during excavation of deep caries. Methods: Human molar teeth with deep occlusal caries were investigated. After obtaining 2-dimensional and 3-dimensional OCT scans using a swept-source OCT system at a 1330-nm center wavelength, RDT was evaluated by image analysis software. Microfocus x-ray computed tomographic (micro-CT) images were obtained from the same cross sections to confirm OCT findings. The smallest RDT values at the visible pulp horn were measured on OCT and micro-CT imaging and compared using the Pearson correlation. Pulpal horns and pulp chamber roof observation under OCT and micro-CT imaging resulted in comparable images that allowed the measurement of coronal dentin thickness. Results: RDT measured by OCT showed optical values range between 140 and 2300 mu m, which corresponded to the range of 92-1524 mu m on micro-CT imaging. A strong correlation was found between the 2 techniques (r=0.96, P<.001). Conclusions: Further analysis indicated linear regression with a slope of 1.54 and no intercept, closely matching the bulk refractive index of dentin. OCT enables visualization of anatomic structures during deep caries excavation. Exposure of the vital dental pulp because of the removal of very thin remaining coronal dentin can be avoided with this novel noninvasive technique.
机译:简介:光学相干断层扫描(OCT)是获取生物结构深入图像的一种非侵入性方式。牙科OCT系统已经可以在椅旁应用。这项体外研究假设,在深龋的开挖过程中,扫频OCT可用于测量牙髓腔顶部残留的牙本质厚度(RDT)。方法:研究具有深咬合龋的人类磨牙。使用扫频OCT系统在1330 nm中心波长获得二维和三维OCT扫描后,通过图像分析软件评估RDT。从相同的横截面获得Microfocus X射线计算机断层扫描(micro-CT)图像,以确认OCT结果。在OCT和micro-CT成像上测量可见果肉角的最小RDT值,并使用Pearson相关性进行比较。在OCT和微型CT成像下观察牙髓角和牙髓腔房顶,可以得到可比较的图像,从而可以测量冠状牙本质的厚度。结果:通过OCT测量的RDT的光学值范围介于140和2300μm之间,对应于微CT成像的92-1524μm范围。发现这两种技术之间有很强的相关性(r = 0.96,P <.001)。结论:进一步的分析表明线性回归具有1.54的斜率并且没有截距,与牙本质的总折射率非常匹配。 OCT可在龋齿深部挖掘过程中可视化解剖结构。用这种新的非侵入性技术可以避免由于去除了非常稀薄的冠状牙本质而暴露出重要的牙髓。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号