Introduction: Recent nickel-titanium manufacturing processes have resulted in an alloy that remains in a twinned martensitic phase at operating temperature. This alloy has been shown to have increased flexibility with added tolerance to cyclic and torsional fatigue. The aim of this study was to assess the effect of repeated simulated clinical use and sterilization on cutting efficiency and flexibility of Hyflex CM rotary files. Methods: Cutting efficiency was determined by measuring the load required to maintain a constant feed rate while instrumenting simulated canals. Flexibility was determined by using a 3-point bending test. Files were autoclaved after each use according to the manufacturer's recommendations. Files were tested through 10 simulated clinical uses. For cutting efficiency, mean data were analyzed by using multiple factor analysis of variance and the Dunnett post hoc test (P < .05). For flexibility, mean data were analyzed by using Levene's Test of Equality of Error and a general linear model (P < .05). Results: No statistically significant decrease in cutting efficiency was noted in groups 2, 5, 6, and 7. A statistically significant decrease in cutting efficiency was noted in groups 3, 4, 8, 9, and 10. No statistically significant decrease in flexibility was noted in groups 2, 3, and 7. A statistically significant decrease in flexibility was noted in groups 4, 5, 6, 8, 9, 10, and 11. Conclusions: Repeated simulated clinical use and sterilization showed no effect on cutting efficiency through 1 use and no effect on flexibility through 2 uses.
机译:简介:近期的镍钛制造工艺已导致合金在工作温度下仍处于孪生马氏体相。该合金已显示出增加的柔韧性,并增加了对循环疲劳和扭转疲劳的耐受性。这项研究的目的是评估重复模拟临床使用和灭菌对Hyflex CM旋转锉刀切割效率和柔韧性的影响。方法:切削效率是通过测量在模拟运河中保持恒定进给速度所需的载荷来确定的。通过使用三点弯曲测试来确定柔韧性。每次使用后,根据制造商的建议对文件进行高压灭菌。通过10种模拟临床用途对文件进行了测试。为了提高切割效率,使用方差多因素分析和Dunnett事后检验分析了平均数据(P <.05)。为了提高灵活性,使用Levene误差均等检验和一般线性模型来分析平均数据(P <.05)。结果:在第2、5、6和7组中,切割效率没有统计学上的显着降低。在第3、4、8、9和10组中,切割效率没有统计学上的显着降低。在第2、3和7组中观察到了弹性。在第4、5、6、8、9、10和11组中观察到了柔韧性的统计学显着降低。结论:反复进行的模拟临床使用和灭菌对切割效率没有影响1次使用不会影响2次使用对灵活性的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号