An improved methodology for investigating the parameters influencing migration resistance of abdominal aortic stent-grafts.

机译:一种改进的方法,用于研究影响腹主动脉支架移植物迁移阻力的参数。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To develop an improved methodology for investigating the parameters influencing stent-graft migration, with particular focus on the limitations of existing methods. METHODS: A physiological silicone rubber abdominal aortic aneurysm (AAA) model for fixation studies was manufactured based on an idealized AAA geometry: the model had a 24-mm neck, a 50-mm aneurysm, 12-mm-diameter legs, a 60 degrees bifurcation angle, and 2-mm-thick walls. The models were authenticated in neck fixation experiments. The displacement force required to migrate stent-grafts in physiological pulsatile flow was tested dynamically in water at 37 degrees C. A commercially available longitudinally rigid stent-graft (AneuRx) and a homemade device with little longitudinal rigidity were studied in a number of different configurations to investigate the effect of neck fixation length and systolic pressure on displacement force. RESULTS: The AneuRx (6.95+/-0.49 to 8.52+/-0.5 N) performed significantly better than the homemade device (2.57+/-0.11 to 4.62+/-0.25 N) in pulsatile flow. The opposite was true in the neck fixation tests because the longitudinal stiffness of the AneuRx was not accounted for. Increasing pressure or decreasing fixation length compromised the fixation of the homemade device. This relationship was not as clear for the AneuRx because decreasing proximal fixation resulted in an increase in iliac fixation, which could assist fixation in this device. CONCLUSION: Assessing the migration resistance of stent-grafts based solely on proximal fixation discriminates against devices that are longitudinally stiff. Current in vivo models may give inaccurate displacement forces due to the high degree of oversizing in these studies. A novel in vitro approach, accounting for longitudinal rigidity and realistic graft oversizing, was developed to determine the resistance of aortic stent-grafts to migration in the period immediately after device implantation.
机译:目的:开发一种改进的方法来研究影响支架移植物迁移的参数,尤其要关注现有方法的局限性。方法:基于理想的AAA几何形状,制作了用于固定研究的生理性硅橡胶腹主动脉瘤(AAA)模型:该模型的颈部直径为24毫米,动脉瘤为50毫米,腿部直径为12毫米,角度为60度分叉角和2毫米厚的壁。该模型在颈部固定实验中得到了验证。在37摄氏度的水中动态测试了在生理脉动流中迁移支架移植物所需的位移力。在许多不同的配置中,研究了市售的纵向刚性支架移植物(AneuRx)和纵向刚度很小的自制装置研究颈部固定长度和收缩压对位移力的影响。结果:AneuRx(6.95 +/- 0.49至8.52 +/- 0.5 N)在脉动流方面的表现明显优于自制装置(2.57 +/- 0.11至4.62 +/- 0.25 N)。在颈部固定测试中,情况恰恰相反,因为未考虑AneuRx的纵向刚度。增大压力或减小固定长度会损害自制设备的固定性。对于AneuRx,这种关系并不十分清楚,因为减少近端固定会导致骨固定的增加,这可能有助于固定该装置。结论:仅基于近端固定来评估支架移植物的迁移阻力会与纵向坚硬的器械区分开。由于这些研究中高度的过大尺寸,当前的体内模型可能会给出不准确的位移力。开发了一种新颖的体外方法,该方法考虑了纵向刚度和现实的移植物过大尺寸,以确定主动脉支架移植物在植入装置后立即对迁移的抵抗力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号