Blood lead misclassification due to defective LeadCare blood lead testing equipment.

机译:由于LeadCare血铅检测设备故障,导致血铅分类错误。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

To the Editor: In May, 2005, a partial recall of Lead-Care~R portable blood lead testing kits was initiated after proficiency testing revealed a negative error (mean, -25%; 95% CI, -15% to -35%) in blood lead concentrations (often called blood lead level in the US) analyzed with this device (1,2). This bias exceeded previously recommended 95% error thresholds established with graphite furnace atomic absorption spectroscopy (3). The recall included 8 lots of defective sensors containing enough material for an estimated 500 000 patient tests distributed between September 2003 and May 2005.
机译:致编辑:2005年5月,在熟练度测试显示负误差(平均值为-25%; 95%CI为-15%至-35%)后,开始部分召回Lead-Care〜R便携式血铅测试套件)的血铅浓度(在美国通常称为血铅水平)(1,2)。该偏差超过了石墨炉原子吸收光谱法(3)建立的先前建议的95%误差阈值。此次召回包括8批有缺陷的传感器,这些传感器包含的材料足够用于估计2003年9月至2005年5月之间分发的50万例患者测试。