首页> 外文期刊>Journal of applied microbiology >Performance of the Sysmex UF1000i system in screening for significant bacteriuria before quantitative culture of aerobic/facultative fast-growth bacteria in a reference hospital
【24h】

Performance of the Sysmex UF1000i system in screening for significant bacteriuria before quantitative culture of aerobic/facultative fast-growth bacteria in a reference hospital

机译:Sysmex UF1000i系统在有氧/兼性速生细菌定量培养之前在参考医院中筛选重要细菌尿的性能

获取原文
获取原文并翻译 | 示例
       

摘要

Objective To evaluate the performance of the Sysmex UF1000i automatic urine screening system in the quantitative culture of fast-growth aerobic/facultative bacteria. Methods and Results A standard procedure was used to recover fast-growth aerobic/facultative micro-organisms in 1225 samples, applying (Sysmex (R)) flow cytometry for parallel bacteria and leucocyte counts. According to the area under the receiver operating characteristic curve, the optimal cut-off values to detect bacteriuria >105 colony forming units (CFU) ml-1 were 690/mu l for bacteria and 38/mu l for leucocytes (sensitivity, 92%; specificity, 65%; positive predictive value [PPV], 39%; and negative predictive value [NPV], 97%). The use of a single cut-off point of 150 bacteria mu l-1 to detect significant bacteriuria of >105 CFU ml-1 or of =104 CFU ml-1 plus leucocyturia obtained similar results (sensitivity, 89%; specificity, 54%; PPV, 31%; and NPV, 96%) and allowed 45.7% of the samples to be rapidly excluded. Conclusions The Sysmex UF1000i system can be adapted for bacteriuria screening by the use of an appropriate cut-off point. Significance and Impact of the Study This screening system significantly reduces the workload and produces very few false positives and negatives.
机译:目的评估Sysmex UF1000i自动尿液筛查系统在快速生长需氧/兼性细菌定量培养中的性能。方法和结果采用标准程序恢复了1225个样品中的快速生长的需氧/兼性微生物,应用(Sysmex(R))流式细胞仪检测平行细菌和白细胞计数。根据接收器工作特性曲线下的面积,细菌> 105集落形成单位(CFU)ml-1的最佳检测截止值对于细菌为690 /μl,对于白细胞为38 /μl(灵敏度为92% ;特异性为65%;阳性预测值[PPV]为39%;阴性预测值[NPV]为97%)。使用150个细菌μl-1的单个截止点检测> 105 CFU ml-1或= 104 CFU ml-1加上白细胞尿症的显着细菌尿液,可获得相似的结果(敏感性为89%;特异性为54% ; PPV为31%; NPV为96%),并允许快速排除45.7%的样品。结论Sysmex UF1000i系统可通过使用适当的临界点进行细菌尿筛查。研究的意义和影响该筛查系统大大减少了工作量,几乎没有假阳性和阴性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号